Preferred Tuning
If the user would like to remain on a QCL during manual tuning, they can click on the header above the QCL
number in the QCL Tuning Range Table to select that QCL as ‘
Preferred
”. Below,
QCL 1
is selected as the
Preferred QCL. The cross-over points for tuning are now ignored and the preferred QCL and tuning will
stay on the Preferred QCL as long as the requested tuning range is within the listed QCL tuning range.
Extended Information is also available by clicking on the
Extended Info
button. This will display
information about each of the QCL channels. It will display the actual QCL current, QCL Voltage, TEC
current, QCL set temperature and actual temperature. To change any settable values, click
Cancel Manual
Tune