Using the Meter

Preparation

 

Select a function and a range for the item to be 

  measured using the rotary switch and “

RAN

” button if 

  necessary. In manual range if the range to be measured 
  is unknown, start with the highest range.

 

In manual range, “

OL

” is displayed when overload 

  occurs.  Choose a higher range. 

 

To perform measurement, first connect the test lead to 

  common (COM) input then connect the probe tip of the 
  Meter to the circuit to be tested.

 

If the battery voltage is less than 2.4V,“     ”is displayed.  

  Please change the batteries.  

DC/AC Voltage Measurement

Use caution when measureing high voltage circuits 
to avoid electrical shock and injury. do not test 
voltages higher than dc/ac 600v.

Warning

 

  or higher measurements.

 

Insert the black test lead or lead with test clip in the 

  “

COM”

 jack.

 

Set the rotary switch to the “

V    

”position.

 

Press the “

FUNC

” button to switch between DC and AC 

  measurements. Manual range can be selected by 
  pressing the “

RAN

” button. Hold “

RAN” 

to go back to 

  Auto range.

 

Connect the probe tip of the meter and tip of the test lead 

  (or test clip) across the voltage source or load.

 

Read the measurement on the display. The polarity of 

  the meter’s “+” probe will be indicated.

Make sure the probe cap is screwed on if making CAT III 

NOTE:

At small voltage ranges the meter may show unsteady 
readings when test leads are not connected to the circuit, 
this is normal because of the sensitivity of the meter. 
When the meter is connected to the circuit, true readings 
will be displayed.

DC/AC Current Measurement

Use caution when measureing high voltage circuits 
to avoid electrical shock and injury. do not test 
voltages higher than dc/ac 600v.

Warning

• Make sure the probe cap is screwed on if making CAT III 
  or higher measurements.
• Insert the black test lead or lead with test clip in the
  “

COM

” jack.

• Set the rotary switch to the “

mA    

”position.

• Press the “FUNC” button to switch between DC and AC 
  measurements. Manual range can be selected by 
  pressing the “RAN” button. Hold “RAN” to go back to 
  Auto range.
• Connect the probe tip of the meter and tip of the test 
  lead (or test clip) across the power source or load.
• Read the measurement on the display. The polarity of 
  the meter’s “+” probe will be indicated.

Содержание DDM350

Страница 1: ...Pen Type Digital Multimeter User s Manual DDM350...

Страница 2: ...stance Measurement 9 Diode Test 9 LIMITED WARRANTY AND LIMITATION OF LIABILITY 1 Safety Symbols 3 Certification 4 Capacitance Measurement 10 Continuity 10 Frequency Duty Cycle 11 Data Hold 12 SPECIFIC...

Страница 3: ...warranty in its entirety is inclusive and no other warranties written or oral are expressed or implied Out of the Box Check the meter and accessories thoroughly before using the meter Contact your lo...

Страница 4: ...dards CE CATIII Complies with U S and Canadian safety standards Never use the meter to measure voltages that might exceed 600V DC AC above earth ground in category III installations Always be careful...

Страница 5: ...old to turn on the work light Hold data reading turn on off backlight NCV Button Hold to activate Non Contact Voltage measurement H Button Switch between frequency duty cycle functions Hz Button Selec...

Страница 6: ...probe tip of the meter and tip of the test lead or test clip across the voltage source or load Read the measurement on the display The polarity of the meter s probe will be indicated Make sure the pr...

Страница 7: ...st clip in the COM jack NOTE The Meter shows the forward voltage drop If the lead connection is reversed OL is displayed Capacitance Measurement To avoid electrical shock and injury power off the circ...

Страница 8: ...me to return to voltage current mode COM To hold a reading press the button and the current reading will be held on the display Press the button again to return to normal reading Backlight Hold the bu...

Страница 9: ...ue Low Battery Indication displayed Fuse Protection FF400mA 600V Storage Temperature 10 50 C 14 F 122 F 70 RH battery removed Max Voltage Between Terminals And Earth Ground 600V DC or AC Auto Ranges a...

Страница 10: ...1 0 reading 1 digits 15 16 AC voltage Resolution 0 1mV Accuracy 0 8 reading 3 digits 400mV 0 001V 0 01V 0 1V 1V 4V 40V 400V 600V 0 8 reading 3 digits Range 1 0 reading 3 digits Input Impedance 10M Ov...

Страница 11: ...sending the meter back for repair include a description of the problems encountered Remove batteries and pack the meter well to avoid damage during delivery Dawson does not cover damage due to delive...

Страница 12: ...d 858 Diamond Bar CA 91765 Phone www DawsonTools com 310 728 6220 Features True RMS LCD Display Auto and Manual Range Auto Power Off Diode Test Audible Continuity Non Contact Voltage Detection Work li...

Отзывы: