Soft Alarms:
The following remaining errors give a 15 second audible chirp only.
System Display
Problem
Resolution
The module indicated is no
longer communicating with
the system.
Ensure all modules are
connected properly to the
system and to each other.
Modules can be tested
individually by plugging in
one at a time and cycling
power to the system.
Replace any module that is
not detected when plugged
directly into the controller.
1. Turbulent flow is present.
2. Air bubbles or solid
particles in the
Flowmeter body/pipe.
3. Internal damage caused
by plugging an AC/DC
wall adapter. that is rated
more than +5VDC or
dropping the device.
1. Make sure there is
proper length of straight
piping on the inlet side
of the Flowmeter. As well
the direction of flow is in
the correct direction.
2. Verify installation is
correct, remove any
air bubbles from water
supply, add filtering to
reduce solid particles
from entering Flowmeter.
3. Replace Flowmeter.
Warning:
After any hard alarm, the home or facility should be disinfected. Follow the
steps under the “System Disinfection” heading.
Boil Water Advisory:
If any failure occurs on a Microlene UV system, the water must
not be used for human consumption until the system is returned to a safe operational
mode. If the water is used for human consumption during this period, the water must
be boiled (minimum 20 minutes at a full boil) prior to consumption.
Temperature Management Devices
Your Microlene system is designed to run continuously to ensure optimal disinfection.
However, during periods when no water is drawn through the system, the energy
from the disinfection process can cause the temperature of the water inside the
chamber to rise. In extreme situations elevated water temperature or the fluctuation
in temperature can lower the output of the UV lamp. In these cases, or if the elevated
water temperature is a nuisance, Microlene recommends the following form of
temperature management device.
Temperature Relief Valve (TRV)
On reaching a higher temperature, the TRV is designed to drain a
small amount of water to allow fresh, cooler water to enter the system.
The TRV works without power and comes complete with
10’ of drain line. The TRV should be used in standard installs where
the ambient temperature regularly exceeds 35 degrees C ambient
temperature. Order PN
DM130033
for 1” ports.