vPad-RW OPERATING MANUAL
Running a Test/Chapter 5
#
Page 43
When all of the temperatures indicated in the Autosequence have been tested, the test will be
considered to be complete, and the app can return to any other test which was in progress. If no
other sensor tests are to be done, the
Probe Test
icon will be removed from the screen. The
Sensor can then be placed in the radiant warmer to perform the
Baby Temperature Control
test. For this test, the Skin Sensor is suspended 10 cm above the center of the mattress. For further
details, see the section on
Baby Temperature Control
.
Figure 43: Sensor Temperature Accuracy - Test Status