![Datex-Ohmeda S/5 M-P Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/datex-ohmeda/s-5-m-p/s-5-m-p_technical-reference-manual-slot_513846016.webp)
Datex-Ohmeda S/5 monitors
Document No. 8001013-1
12
Pressure sensor connector (X4)
Pin No
board
Pin No
fr.panel
Signal
Notes
1
0
PCABEL
5V if cable not connected
2
4
PCURRENT
Pulsed supply to sensor
3
3,9
P3-
Signal from the sensor
4
1
P3+
Signal from the sensor
5
2
fGND
Floating GND
X4
1
5
When the board is used in the M-PT module, there are the following connectors connected on the
board in addition to X1 and X4.
Signal out connector (X5)
Pin No
board
Pin No
fr.panel
Signal
Notes
1
-
ECG OUT
ECG to (e.g.) IABP
2
-
N/C
Not connected
3
4
GND
+5VDCin GND
4
1
ECG OUT
ECG to (e.g.) defib.
5
-
N/C
Not connected
6
3
ECG OUT
ECG to (e.g.) IABP
7
2
P OUT
P3 to (e.g.) IABP
8
GND
GND
+5VDCin GND
X5
1
8
Connector for temperature probes (X8)
Pin No
Signal
Notes
1
T3
-
2
T4
-
3
fGND
Floating GND
4
CON
Low if PT
5
fGND
Floating GND
6
GND
Floating GND
X8
1
6