![Datex-Ohmeda S/5 B-ARK Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/datex-ohmeda/s-5-b-ark/s-5-b-ark_technical-reference-manual-slot_3786203013.webp)
S/5 Anesthesia Record Keeping Keyboard and Barcode Reader
Document No. 8001019-3
9
3.2
Service check
These instructions include complete procedures for a service check. The service check is
recommended to be performed after any service repair. However, the service check procedures can
also be used for determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form (
Appendix A
) which should be filled in when performing the
procedures.
The mark
?
in the instructions means that the check form should be signed after performing
the procedure.
The procedures are designed for monitors with software of level 00. However, most of the
procedures also apply to monitors with older monitor software.
3.2.1
Recommended tools
Tool
Order No.
Notes
B-ARK
with AM
Screwdriver
1.
Disconnect the interface cable from the keyboard for the deioRecorder for Anesthesia, K-
ARKB and check the cable:
−
the connector pins are clean and straight and at about the same height
−
the locking screws are intact
−
the cable is intact
Leave the cable disconnected.
?
2.
Detach the bottom cover and check internal parts:
−
all screws are tightened properly
−
the block screws for the interface cable are in place and are tightened properly
−
the block screw threads are intact
−
the interface cable connector is clean and intact
−
all internal cables are connected properly
−
all IC’s that are on sockets are attached properly
−
there are no loose objects inside
Reattach the bottom cover, reconnect and lock the interface cable to the keyboard for the
deioRecorder for Anesthesia, K-ARKB.
?
Содержание S/5 B-ARK
Страница 2: ......
Страница 4: ...Datex Ohmeda S 5 monitors Document No 8001019 3 ii SERVICE CHECK FORM A 1...
Страница 26: ...Datex Ohmeda S 5 monitors Document No 8001019 3 22...
Страница 27: ...APPENDIX A Service check form K ARKB Document No 8001019 3 23 APPENDIX A...
Страница 28: ...Datex Ohmeda S 5 monitors Document No 8001019 3 24...
Страница 30: ...Datex Ohmeda S 5 monitors Document No 8001019 3...