NeuroMuscular Transmission Module, M-NMT
Document No. 800 1018-1
5
•
microprocessor for the stimulation and measuring control, and for counting the measuring
results
•
serial communication
The serial bus speed is 500 kbps and the bus itself is half duplex, i.e. data can be transferred in
both directions but only one way at a time.
Serial communication
Figure 3
Serial communication and opto isolation
Stimulator
The constant current stimulator generates pulses whose amplitude is independent of the load. The
main components of the stimulator are a transformer, a capacitor and a transistor. The transformer
produces a high voltage which charges the capacitor and the transistor adjusts the pulse width and
amplitude of the current.
Patient
Constant current
generator
Power supply
Current
adjustment
Current
control
Continuous
stimulation
congestion
80C196KD
PWM
HSO
Figure 4
Stimulator block diagram