D-Tec Vaporizer
Page 9
O & M Manual Part No. 1107-0622-000
August 2001
DURACELL
D U R A
C E L L
m n 62
3 r
4.0
Vaporizer Preparation
4.1
Fitting The Battery And The Mains Lead
w
Caution: Do not support the vaporizer by holding the control dial and/or locking lever. Hold the main
body of the vaporizer with both hands and keep the vaporizer in an upright position. Failure to do so
may cause the vaporizer to malfunction.
w
Caution: Only a Duracell 1604 or VARTA Energy 2000, 9 Volt battery must be installed in the
vaporizer. If any other battery is installed, it may damage the vaporizer.
w
Warning: The battery terminals must be firmly clipped onto the battery to help prevent a possible
disconnection when the vaporizer is moved.
w
Warning: Only a Datex-Ohmeda mains lead must be used to connect the vaporizer to the electrical
supply.
w
Warning: When routing the mains lead to the electrical supply, ensure that it does not interfere with
the correct functioning of other equipment.
The new battery and the mains lead supplied with the vaporizer must be fitted during initial installation
as detailed in Instructions 1, 2 and 3, therefore Instructions 1, 2 and 3 can be ignored if a battery and
mains lead are already fitted.
1.
Invert the vaporizer, unscrew the battery cover securing screw and remove the battery cover from
the base of the vaporizer as illustrated on Fig. 10.
2.
Clip the battery terminals firmly onto the new battery observing the correct polarity. Insert the
battery into the vaporizer, fit the battery cover to the base of the vaporizer and tighten the securing
screw.
Fig. 10 Fitting The Battery
BATTERY
TERMINALS
BATTERY COVER
SECURING SCREW
VAPORIZER
INVERTED
DURACELL 1604 OR VARTA
ENERGY 2000 BATTERY
AA.43.044
Содержание D-Tec
Страница 1: ...D Tec Vaporizer Operation and Maintenance Manual ...
Страница 3: ...D Tec Vaporizer ...