Datex-Ohmeda ADU
C-80
Document No. 8502855
Leak tests for service purposes
These tests are for detecting leaks in specific areas of the ventilator and the bellows block. The test in
“Ventilator test (Test 1)” on page C-80, includes expiration and inspiration valves, flow measurement
unit, drive gas cylinder, bottle pressure line from the bellows block bottom plate to the chamber
pressure sensor and zeroing valve as well as PEEP support line and peep support valve in the same
way.
The test in “Test for circuit and monitor pressure sensor lines (Test 2)” on page C-80, involves the
pressure lines to the circuit and the monitor pressure sensors and their zeroing valves.
The purpose of the test in “Test for Bellows block and bottom plate leakage (Test 3)” on page C-81, is
to identify leakages in the bellows block and the bottom plate of the ADU for service purposes. The
following procedure is for testing the possible points of leakage in Auto ventilation mode.
Ventilator test (Test 1)
1.
The ADU in power On
2.
The gas inlets connected to the central outlets
3.
Enter Pneumatics Control on the Ventilator diagnostic page and Operate the expiration valves
(EXP. and PEEP on the diagnostic page). Close them both.
4.
Remove the bellows block.
5.
Connect a pressure manometer (cuff pump) to the driving gas hole in the bellows block bottom
plate with a suitable connector. See the figure below.
6.
Pressurize the volume with the cuff manometer up to 30 mmHg.
Figure 46
The channels in the Bellows Block bottom plate
7.
Wait for a minute. A pressure drop of 2 mmHg in 10 seconds at 30 mmHg pressure corresponds
approximately 10 ml/min. leakage. Pressure can also be observed from the output of bottle
pressure sensor on the diagnostics page of the ventilator.
Test for circuit and monitor pressure sensor lines (Test 2)
1.
Remove the bottom plate gasket.
2.
Connect the cuff manometer to the circuit and to the monitor pressure hole of the bellows
bottom plate. See the figure above.
3.
Pressurize the circuit lines to 30 mmHg.
NOTE: Do not apply too high pressure to the lines as it may damage the pressure sensors.
4.
A possible leakage can be detected by the pressure drop on the manometer or by the circuit
and monitor sensor values on the ventilator service menu.
Scavenging
Driving gas
PEEP support
bottle pressure
Circuit and monitor pressure
Содержание Anesthesia Delivery Unit S/5
Страница 2: ......
Страница 14: ......
Страница 36: ...Datex Ohmeda ADU 3 14 Document No 8502852 This page is intentionally left blank ...
Страница 62: ...Maintenance Document No 8502852 5 4 Figure 1 Service menu overview ...
Страница 76: ...Datex Ohmeda ADU 5 18 Document No 8502852 Figure 24 Transformer unit Figure 25 Transformer unit 2 1 4 6 ...
Страница 80: ...Datex Ohmeda ADU 5 22 Document No 8502852 Figure 30 Compact Circuit A CC1 exploded view 5 1 2 3 4 ...
Страница 88: ...Maintenance 5 30 Document No 8502852 Figure 38 Maintenance parts ...
Страница 114: ...Datex Ohmeda ADU 5 56 Document No 8502852 This page is intentionally left blank ...
Страница 132: ...Forms and charts 7 8 Document No 8502852 Pneumatic chart of the Electronic Ventilator A EV1 ...
Страница 133: ...Forms and charts Document No 8502852 7 9 Pneumatic chart of the Fresh Gas Control Unit A FGC1 ...
Страница 135: ...Forms and charts Document No 8502852 7 11 Pneumatic chart of Table frame without A FGFI ...
Страница 136: ...Datex Ohmeda ADU 7 12 Document No 8502852 This page is intentionally left blank ...
Страница 137: ...Document No 8502852 7 13 Pneumatic chart of A WGU frame without A FGFI ...
Страница 138: ...7 14 Document No 8502852 Pneumatic chart of A WGU frame with A FGFI ...
Страница 139: ...Document No 8502852 7 15 Pneumatic chart of Table frame without A FGFI ...
Страница 140: ...7 16 Document No 8502852 Pneumatic diagram of table frame with A FGFI ...
Страница 146: ...Datex Ohmeda ADU 8 6 Document No 8502852 This page is intentionally left blank ...
Страница 147: ...Technical bulletins Document No 8502852 9 1 9 TECHNICAL BULLETINS ...
Страница 148: ......
Страница 152: ...Datex Ohmeda ADU 10 4 Document No 8502852 This page is intentionally left blank ...
Страница 194: ...Datex Ohmeda ADU A 42 Document No 8502853 This page is intentionally left blank ...
Страница 198: ...Datex Ohmeda ADU A 46 Document No 8502853 This page is intentionally left blank ...
Страница 211: ...Spare parts A AUF Document No 8502853 A 59 Figure 53 Exploded view of the gas inlet block and the gas labels 4 1 2 3 ...
Страница 229: ...Technical description A FGC1 Document No 8502854 B 3 Figure 3 Technical principles of the A FGC1 ...
Страница 249: ...Technical description A FGC1 Document No 8502854 B 23 This page is intentionally left blank ...
Страница 257: ...Technical description A FGC1 Document No 8502854 B 31 Figure 25 CPU Board block diagram ...
Страница 266: ...Datex Ohmeda ADU B 40 Document No 8502854 This page is intentionally left blank ...
Страница 304: ...Datex Ohmeda ADU B 78 Document No 8502854 This page is intentionally left blank ...
Страница 307: ...Spare parts A FGC1 Document No 8502854 B 81 Figure 60 Fresh Gas Control Unit A FGC1 1 2 4 3 6 5 7 8 10 11 12 ...
Страница 311: ...Spare parts A FGC1 Document No 8502854 B 85 Figure 64 Parts inside the A FGC1 1 2 3 4 5 6 7 8 10 ii iii i iv 9 ...
Страница 319: ...Spare parts A FGC1 Document No 8502854 B 93 Figure 71 Gas connection unit 1 2 3 4 5 6 7 8 9 10 11 ...
Страница 326: ...Datex Ohmeda ADU B 100 Document No 8502854 This page is intentionally left blank ...
Страница 360: ...Datex Ohmeda ADU C 34 Document No 8502855 This page is intentionally left blank ...
Страница 362: ...Datex Ohmeda ADU C 36 Document No 8502855 The ventilator calibrating tree Figure 27 The ventilator calibrating tree ...
Страница 400: ...Datex Ohmeda ADU C 74 Document No 8502855 Ventilator diagnostic page layout Figure 42 Ventilator diagnostic page ...
Страница 408: ...Datex Ohmeda ADU C 82 Document No 8502855 This page is intentionally left blank ...
Страница 417: ...Spare parts A EV1 Document No 8502855 C 91 This page is intentionally left blank ...
Страница 421: ...Spare parts A EV1 Document No 8502855 C 95 This page is intentionally left blank ...
Страница 432: ...Datex Ohmeda ADU D 2 Document No 8502856 Figure 1 Principle of the Central Electronics ...
Страница 477: ...Troubleshooting A ELEC Document No 8502856 D 47 Trouble shooting PCMCIA Card Figure 57 Troubleshooting PCMCIA Card ...
Страница 480: ...Datex Ohmeda ADU D 50 Document No 8502856 This page is intentionally left blank ...
Страница 483: ...Spare parts A ELEC Document No 8502856 D 53 Figure 60 11 12 1 2 4 3 5 6 3 8 9 13 14 15 16 10 ...
Страница 486: ...Datex Ohmeda ADU D 56 Document No 8502856 Figure 63 Transformer unit Figure 64 Transformer unit 2 1 4 ...
Страница 487: ...Spare parts A ELEC Document No 8502856 D 57 Figure 65 Transformer unit 5 3 ...