Aespire View
10-32
M1132382
System components
Integral
This anesthesia system contains the following integral components,
monitoring devices, alarm systems, and protection devices that
comply with European, international, and national standards:
•
Breathing system pressure-measuring device.
•
Airway pressure-limitation device.
•
Exhaled-volume monitor.
•
Breathing system integrity alarm.
•
Breathing system continuing-pressure alarm.
•
O2 monitor.
•
Anesthesia ventilator.
•
Breathing system.
Not integral
These devices are not integral to this anesthesia system:
•
CO2 monitor.
•
Anesthetic agent monitor.
WARNING
European, international, and national standards require
the following monitoring be used with this system:
•
Exhaled volume monitoring.
•
O2 monitoring.
•
CO2 monitoring.
•
Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
When adding devices to the anesthesia systems, follow the
installation instructions provided by the device manufacturer.
Whoever adds individual devices to the anesthesia system shall
provide instructions on how to enable the individual devices (for
example, a preoperative checklist).
Содержание Aespire View
Страница 1: ...Aespire View User s Reference Manual Software Revision 6 X...
Страница 16: ...Aespire View 1 8 M1132382...
Страница 46: ...Aespire View 3 16 M1132382...
Страница 50: ...Aespire View 4 4 M1132382...
Страница 88: ...Aespire View 7 8 M1132382...
Страница 112: ...Aespire View 9 10 M1132382...
Страница 114: ...Aespire View 10 2 M1132382 System pneumatic circuits Figure 10 1 Pneumatic circuit diagram AC 20 001...
Страница 118: ...Aespire View 10 6 M1132382 Electrical block diagram Figure 10 2 Electrical block diagram AC 20 008...
Страница 137: ...10 Specifications and Theory of Operation M1132382 10 25 Figure 10 9 Gas composition related errors AB 74 027...
Страница 148: ...Aespire View I 4 M1132382...