9 Schematics and Diagrams
1006-0836-000 02/03
9-5
Figure 9-4 • Aespire 7100 anesthesia machine pneumatic diagram
PT
PT
PT
NC
P
O2 Cyl (opt)
Patient
Atmosphere
Bag
Ambient
Control Bleed to Ambient
1.0 LPM @ 3.0 cm H2O
if continuous (rate dependent)
Bag/Vent
5.4 psi
P
P
Aux O2
0-10 LPM
(opt)
O2 Flush
O2 P-LINE
AIR P-LINE (opt)
N2O P-LINE (opt, Std US)
110 psi
Relief
N2O Cyl (opt)
O2 Cyl (opt)
25 psi
@ 15 LPM
Venturi
Exp
. Flo
w
T
rans
.
Exhalation Valve
(2.0 cm H2O bias)
Free Breathing
Check Valve
Expiratory
Flow Sensor
0-120 LPM Flow
Mechanical Overpressure
Valve (110 cm H2O)
P
Air Cyl (opt)
Vent
Drive
0-10 LPM Drive Gas
0-10 LPM Patient and Fresh Gas
0-20 LPM Total Typical Flow
35 psi
Gas Monitor
Inspiratory
Flow
Control
Valve
Popoff
Valve
O2
Sensor
(opt)
110 psi
Relief
110 psi
Relief
Insp
. Flo
w
T
rans
.
P
200 mL Reservoir
APL valve
0-70 cm H2O
Absorber
Dr
ain
Selectatec Manifold
VAP
VAP
5.5 psi Pressure
Relief Valve
Flowmeter Module
Link 25
Link 25
30 psi
System Switch
(Dual Flowtubes Optional)
PEEP Safety Valve
Proportional PEEP Valve
Supply Pressure Switch
P
Gauge
Gauge
Gauge
P
Gauge
P
Gauge
Gauge
Gauge
Negative Pressure
relief valve
10 cm H2O
Ventilator Monitoring Board
Airway
Pressure
Gauge
NO
NC
NO
NC
Pneumatic
Engine
Board
ACGO
Selector
Valve
Balance
Regulator
To Scavenging
NO
O2
Flush
Switch
Inspiratory
Flow Sensor
O2
Supply
Switch
Air (opt)
30 psi
ACGO
Venturi
AB.74.029