DI-1110 Hardware Manual
Controls, Indicators, and Connections
13
the designated Rate channel. Click on the Channel 10 channel box to enable/disable Rates. The Channel 10 channel
box will display an “R” when enabled.
Connect any TTL pulse stream (25 VDC max) to the D2 Rate channel on the DI-1110.
Select Channel 10 in the channel display, then select
Edit
>
Channel Settings
in the menu to specify a full scale
range for the Rate channel. Ranges of 10, 20, 50, 100, 200, 500, 1000, 2000, 5000, 10000, 20000, and 50000 Hz are
available. Choose a range that comes closest to your expected full scale rate or frequency. For example, if you need to
measure the output of a flow meter that can measure a maximum flow rate of 100 gallons per minute, and the output
frequency at that rate is 100 Hz, you would choose the 100 Hz range setting. A range setting of 200 Hz full scale sac-
Signal -
www.dataq.com
+
-
+
-
+
-
+
-
D5
D4
D6
GnD
+5V
Power
Remote
Active
-
+
Ch1
-
+
Ch2
-
+
Ch3
-
+
Ch4