background image

DI–1000TC User Manual

Getting Started

10

Installing Multiple Devices in a Daisy Chain

Up to 8 DI-1000TC instruments (64 channels max) may be daisy-chained together to form a distributed synchronous 
network.

1.

Connect the first instrument in the chain to your adapter (DI-1000-USB or DI-1000-232) and install it using the 
appropriate installation instructions. This is the “Master” instrument.

2.

Run the DI-1000 Configuration Software found in the Windows Program Group specified during installation. 
The default is Start > Programs > WinDaq > DI-1000 Config.

3.

Click on Configure > Setup in the DI-1000 Config menu.

4.

Enter “1” in the text boxes for Address, Previous Device, and Master.

5.

If you are daisy-chaining one to four units, select a Baud Rate of 9600 (default) for 1-30 channels or 19200 for 
more than 30 channels. If you are daisy-chaining five to eight units, select a Baud Rate of 19200.

6.

Click on 

Write

 to configure the instrument.

7.

Disconnect the “Master” instrument from the adapter.

8.

Connect and power the next instrument in the chain.

9.

Enter “2” for Address. Enter “1” for Previous Device, and Master. This device is “Slave 1.” Enter the same Baud 
Rate entered in Step 5. For more information on USB ID addresses see “Setting up Addresses for Your Devices” 
on page 13.

10. Repeat steps 7 and 8 for each device incrementing Address and Previous Device by one for each subsequent 

instrument. Each unit must be powered down after changing the device address or baud rate. Master should 
always be “1” and the Baud Rate should be the same for each device.

11. After the setup is complete for all devices in the chain, connect the “Master” unit to the adapter.

12. Insert a standard CAT-5 cable to one of the RS-485/422 ports on the back of the “Master” instrument. It does not 

matter which port you insert the cable into.

13. Connect the other end of the CAT-5 cable to one of the RS-485/422 ports on the back of the “Slave 1” instrument. 

It does not matter which port you insert the cable into.

14. Connect each instrument in order by Address (assigned in step 9) in the same manner.

15. Power each instrument using one of the Power jacks on the rear of each instrument. Power may be daisy-chained 

in the same manner as the CAT-5 cables (four instruments max for each power supply).

16. Once all units have been connected and powered you must change the Target in the Shortcut Properties in 

WinDaq Software to reflect the number of “Slave” devices in your network.

a.   Right-click on the WinDaq Lite Data Acq DI-1000TC program in the Windows startup menu.  Default is Start > 

Programs > WinDaq > WinDaq Lite Data Acq DI-1000TC.

b.   Click on the Shortcut tab.

c.   Change “-N0” to “-Nx” at the end of the Target text box where x is the number of “Slave” units installed in the 

daisy chain (total number of instruments minus 1).

Содержание DI-1000TC

Страница 1: ...herwise without expressed written authorization from the com pany The distribution of this material outside the company may occur only as authorized by the company in writing DATAQ Instruments hardwar...

Страница 2: ......

Страница 3: ...products that have been repaired or altered by persons other than DATAQ Instru ments employees or products that have been subjected to misuse neglect improper installation or accident DATAQ Instrument...

Страница 4: ......

Страница 5: ...er to the DI 1000TC Instrument 5 Connecting the DI 1000TC Instrument to your Computer 5 RS 422 Interface Port Pin Out 6 Adapters 6 DI 1000TC Instrument Controls and Indicators 6 DI 1000TC Instrument F...

Страница 6: ......

Страница 7: ...1000TC s 1000VDC and Peak AC input to output and channel to channel isolation allows grounded TC measurements High Accuracy and Resolution Design Allows sharpened decisions with an overall accuracy o...

Страница 8: ......

Страница 9: ...n configuration Cold Junction Compensation Range 40 to 80 C Linearization Accomplished transparently by the software driver to deliver scaled tem perature values to the host program Input Impedance 1M...

Страница 10: ...in the field Supported Baud Rates 4800 9600 default 19200 38400 56800 115200 Data Bits 8 Stop Bits 1 Parity None Protocol ModBus ASCII Connector RJ 45 Pinout is as follows 1 Receive 2 Receive 3 XMIT 6...

Страница 11: ...wer to your DI 1000 Series instrument 1 Plug the DC power plug end of the power adapter cable into the DC power jack labeled POWER on the rear panel of the instrument There are two identical jacks eit...

Страница 12: ...000TC Instrument Controls and Indicators The following describes the user accessible items on the DI 1000TC Series instruments DI 1000TC Instrument Front Panel The DI 1000TC units are provided in 4 or...

Страница 13: ...if you would like to provide your own power source Installing the Software The software required to display and record data from the DI 1000TC can be easily installed directly from the WINDAQ Resource...

Страница 14: ...specify a destination or group window in the Start Menu for WinDaq software icons It is recommended that you accept the default 15 In the Installation Option dialog box specify whether you want all u...

Страница 15: ...are disconnected Click OK to continue 12 Connect the DI 1000 USB adapter to your computer s USB Port using the supplied cable 13 Connect the DI 1000TC to the DI 1000 USB adapter using a patch cable 1...

Страница 16: ...or each device incrementing Address and Previous Device by one for each subsequent instrument Each unit must be powered down after changing the device address or baud rate Master should always be 1 an...

Страница 17: ...DI 1000TC User Manual Getting Started 11 d Click OK to close the Properties dialog box Your distributed synchronous data acquisition system for thermocouple measurements is installed and ready to run...

Страница 18: ......

Страница 19: ...ice Note that they all must have the same Baud Rate and the addresses must be synchronized see below Note Each device must be connected and configured separately before daisy chaining Changes take eff...

Страница 20: ...ens for the module having a base address equal to the value programmed into it as Previous Address to send its data to the network After it detects that the module with Previous Address has sent its d...

Страница 21: ...sample rate per channel For example if you enter 5 as your sample rate per channel and you have all 8 channels enabled your current sample rate samples per sec ond would be 40 samples per channel tim...

Страница 22: ...want your data to appear in F or C C is the default You can toggle between F and C by select ing and deselecting the Fahrenheit check box Once you have selected your thermocouple type that thermocou p...

Страница 23: ...d be selected across from the 2 in the Addr column on the right of the window These channels are labeled 9 16 and they correspond with 1 8 on your device For example if you wanted to select channels 1...

Страница 24: ...han one device The section after the colon is an annota tion for the channel you have selected For example you may be comparing multiple engines in different locations You may want to label your first...

Страница 25: ......

Страница 26: ...41 Springside Drive Akron Ohio 44333 Telephone 330 668 1444 Fax 330 666 5434 E mail support dataq com Direct Product Links click on text to jump to page Data Acquisition Data Logger Chart Recorder The...

Отзывы: