background image

9

Remove the platter and 

push the wire lift handle 

to the upright position in 

preparation for sealing the 

calibration door.

10

Fully close

 the calibration 

door and apply the 

calibration seal in accor-

dance with local require-

ments. The door is fully 

closed when the holes in 

the door and the metal 

tab are aligned. (See 

Sealing Notes.)

No portion of the seal, 

whether wire, paper 

label or other method, 

should be in contact 

with the weigh frame 

(shown in 

green

) or 

scale movement will be 

compomised.

Lift Handle

Sealing Notes

Conditions to Avoid:

Wire loop too long (not snug)

Calibration door not closed

Seal in contact with weigh frame

CALL DATALOGIC TECHNICAL SUPPORT:

 1-800-547-2507

VIEW THE CALIBRATION VIDEO AT: http://www.youtube.com/watch?v=wHYjcTdd9Gk

IF YOU ENCOUNTER PROBLEMS OR HAVE QUESTIONS ABOUT

ANY PART OF THE CALIBRATION PROCESS:

                                                   

820061014 (Rev. A)

October 2013