9
Remove the platter and
push the wire lift handle
to the upright position in
preparation for sealing the
calibration door.
10
Fully close
the calibration
door and apply the
calibration seal in accor-
dance with local require-
ments. The door is fully
closed when the holes in
the door and the metal
tab are aligned. (See
Sealing Notes.)
No portion of the seal,
whether wire, paper
label or other method,
should be in contact
with the weigh frame
(shown in
green
) or
scale movement will be
compomised.
Lift Handle
Sealing Notes
Conditions to Avoid:
Wire loop too long (not snug)
Calibration door not closed
Seal in contact with weigh frame
CALL DATALOGIC TECHNICAL SUPPORT:
1-800-547-2507
VIEW THE CALIBRATION VIDEO AT: http://www.youtube.com/watch?v=wHYjcTdd9Gk
IF YOU ENCOUNTER PROBLEMS OR HAVE QUESTIONS ABOUT
ANY PART OF THE CALIBRATION PROCESS: