 83/8400 Calibration Process Quick Reference Sheet

1

2

3

4

5

6

7

8

Ensure that the Scanner- 

Scale is powered-on (the 

green top LED and orange 

>0< button will be lit or 

flashing).

Return the wire lift handle 

to its down position and 

replace the Top Platter. (If 

the Top Platter does not 

seat correctly, recheck to 

see that the lift handle is 

fully retracted.) 

After ensuring the platter 

is fully seated, press & 

release the flashing >0< 

button 

once

. Wait for the 

scanner to sound two 

beeps. (The >0< button 

will continue to flash.)

Place 20 lbs. of certified 

weights on the platter and 

press & release the >0< 

button 

once

. The scanner 

will emit a trill sound upon 

registering the weight.

Remove the weights from 

the platter and press & 

release the >0< button 

once

 

to complete the calibration 

sequence. The scanner will 

reset and the >0< button 

will light steadily. 

Remove the Top Platter 

from the scanner-scale, 

and locate the switch 

door.       

Open the calibration door 

cover by pushing back 

towards the optics 

chassis. (It may be 

necessary to push the 

wire lift handle into the 

upright position to access 

the cover.)       

Use your finger to reach 

inside the calibration bay, 

locate the small plunger 

switch and press & 

release it 

once

. The 

scanner will issue seven 

rapid beeps and the >0<  

button will begin to flash.