Datacolor Conditioner Operation Manual / DCCOPMN_RP1 / Rev 3.0 / 7th August 07
page 15 of 51
3.
Scope of delivery, transportation, storage and installation
3.1
Unpacking, and checking equipment and scope of delivery
After having unpacked, please check the unit and its optional accessories, if any, based on the delivery
note for completeness and for transportation damage. If transportation damage has occurred,
immediately inform the carrier.
The final tests of the manufacturer might cause traces of the racks at the inner surfaces. This has no
impact on the function and performance of the unit.
Please remove any transportation protection devices and adhesives in / on the unit and at the doors and
take out the operation manuals and accessory equipment.
CAUTION
Sliding or tilting of the unit.
Damage of the unit.
∅
Do NOT lift or transport the unit using either the door handle or the door or at the lower
housing.
¾
Lift unit near the 4 unit feet from the pallet by aid of 4 persons.
If necessary to send back the unit, please use the original packing and respect the advice for safe lifting
and transportation (chap. 3.2).
For disposal of the transport packing, see chap. 12.1.
3.2
Advice for safe lifting and transportation
Following operation, the unit must be drained off for transport. To pump off the water from the humidifying
system, close the water supply, switch on humidity switch (5), and switch on and off the unit twice
allowing it each time to run about 2 minutes. Respect the advice for temporal decommissioning (chap.
12.2).
CAUTION
Sliding or tilting of the unit.
Damage of the unit.
¾
Transport the unit only in its original packaging.
¾
Secure the sample conditioning cabinet with transport straps for transport.
∅
Do NOT lift or transport the unit using either the door handle or the door or at the lower
housing.
¾
Lift unit near the 4 unit feet by aid of 4 persons.
•
Permissible ambient temperature range during transport: -10°C to +60°C.
3.3 Storage
Intermediate storage of the unit is possible in a closed and dry room. Respect the advice for temporal
decommissioning (chap. 12.2).
•
Permissible ambient temperature range during storage: -10 °C to +60 °C.
•
Permissible ambient humidity during storage: max. 70 % RH, non-condensing
Store units only with the humidifying system drained, and ensure the interior is dry. To pump off the water
from the humidifying system, close the water supply, switch on humidity switch (5), and switch on and off
the unit twice allowing it each time to run about 2 minutes.