Datacolor CONDITIONER™ (MB2) 06/2017
page 21/126
3.4
Location of installation and ambient conditions
Set up the cabinet on a flat, even surface, and in a well-ventilated, dry location and align it using a spirit
level. The site of installation must be capable of supporting the cabinet’s weight (see technical data, chap.
23.4). The cabinets are designed for setting up inside a building (indoor use).
CAUTION
Danger of overheating.
Damage to the cabinet.
∅
Do NOT set up cabinet in non-ventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
•
Permissible ambient temperature range during operation
: +18 °C /
64.4 °F
to +32 °C /
89.6 °F
. At
elevated ambient temperature values, fluctuations in temperature can occur.
The ambient temperature should not be substantially higher than the indicated ambient tem-
perature of +22 °C +/- 3 °C /
71.6 °F ± 5.4 °F
to which the specified technical data relate. De-
viations from the indicated data are possible for other ambient conditions. Lower values of the
temperature range indicated in the technical data are valid at an ambient temperature of max.
25 °C
/ 77 °F
.
With each degree of ambient temperature >25 °C /
77 °F
, the refrigeration power decreases
by 1.5 K.
•
Permissible ambient humidity
: 70 % r.H. max., non-condensing
When operating the cabinet at temperature set-points below ambient temperature, high ambient humidity
may lead to condensation on the cabinet.
•
Installation height: max. 2000 m /
6562 ft.
above sea level.
A water tap (1 bar to 10 bar) is necessary for the installation of the humidification system (chap. 4.3). If no
suitable in-house water connection is available, you can manually supply water by filling the freshwater
can (option, chap. 4.5).
To avoid any possible water damage, provide a floor drain at the location of the device. Select
a suitable installation site to avoid any consequential damage by splashing water.
When placing several cabinets of the same size side by side, maintain a minimum distance of 250 mm /
9.84 in
between each cabinet. Wall distances: rear 100 mm /
3.9 in
, sides 160 mm /
6.29 in
. Spacing
above the cabinet of at least 100 mm /
3.9 in
must also be accounted for.
CAUTION
Danger by stacking.
Damage to the cabinets.
∅
Do NOT place the cabinets on top of each other.
To completely separate the cabinet from the power supply, you must disconnect the power plug. Install
the cabinet in a way that the power plug is easily accessible and can be easily pulled in case of danger.
For the user there is no risk of temporary overvoltages
in the sense of EN 61010-1:2010.
With an increased amount of dust in the ambient air, clean the condenser fan (by suction or blowing)
several times a year.
Avoid any conductive dust in the ambiance according to the cabinet layout complying with pollution de-
gree 2 (IEC 61010-1).