Connecting Current Loop Inputs to RTD Channels
RTD channels on the DT9872 and DT9874 instruments have an input range of ±1.25 V.
Therefore, you can use a 1
shunt resistor to measure ±1.25 A. Similarly, you can use a 0.1
shunt resistor to measure ±12.5 A or a 10
shunt resistor to measure ±0.125 A.
shows how to wire your signals to measure a current loop. In this example, the input
is referenced to ±60 V.
Note:
On the standard DT9874 instrument, channels 16 to 31 correspond to the RTD input
channels.
Figure 22: Connecting Current Loop Inputs to RTD Channels
1
shunt resistor
Use a 1
shunt resistor to convert current to voltage: 1
= 1.25 A = 1.25 V.
60 V
+
–
RTD Channel
– Sense
+Sense
1
2
4
3
Shield
Содержание DT9871
Страница 4: ......
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Hardware Features 14 Supported Software 19 Accessories 21 Getting Started Procedure 22 ...
Страница 23: ...Part 1 Getting Started ...
Страница 24: ......
Страница 50: ...Chapter 4 50 Figure 16 2 Wire RTD Connection RTD Channel RTD 425 A Current Sense Sense Return RL RL 1 2 4 3 ...
Страница 80: ...Chapter 5 80 ...
Страница 81: ...Part 2 Using Your Instrument ...
Страница 82: ......
Страница 83: ...83 6 Principles of Operation Block Diagrams 84 Analog Input Features 89 Digital I O Features 100 ...
Страница 102: ...Chapter 6 102 ...
Страница 103: ...103 7 Troubleshooting General Checklist 104 Technical Support 106 If Your Instrument Needs Factory Service 107 ...
Страница 108: ...Chapter 7 108 ...
Страница 129: ...129 B Connector Pin Assignments ...