Chapter 5
54
Digital I/O Features
The DT9824 module provides 8 digital input lines and 8 digital output lines that you can use
to control external equipment, including solid-state or mechanical relays.
This section describes the following digital I/O features of the DT9824 modules:
• Digital input lines, described below
• Digital output lines, described on
• Channel-to-channel isolation, described on
• Resolution, described on
• Operation modes, described on
Digital Input Lines
The DT9824 module features eight, isolated, digital input lines.
Digital inputs operate from +3 to +28 V DC, with a switching time of 2 ms maximum.
shows the digital input circuitry; a 2.2 k
Ω
resistor is used in series with the LED in the
opto-isolator input.
Figure 15: Digital Input Circuitry
A digital line is high (switch is closed) if its value is 1; a digital line is low (switch is open) if its
value is 0.
Digital Output Lines
The DT9824 module features eight, latched and isolated digital output lines. The outputs are
solid-state relays that operate at ±30 V and 400 mA peak (AC or DC). Switching time is 2 ms
maximum.
shows the digital output circuitry.
Digital Input
+
-
2.2K
Содержание DT9824
Страница 1: ...DT9824 UM 23782 G User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Features 14 Supported Software 16 Getting Started Procedure 18 ...
Страница 19: ...Part 1 Getting Started ...
Страница 20: ......
Страница 21: ...21 2 Installing the Module Unpacking 23 Attaching Modules to the Computer 24 Configuring the Driver 27 ...
Страница 28: ...Chapter 2 28 ...
Страница 44: ...Chapter 4 44 ...
Страница 45: ...45 Part 2 Using Your Module ...
Страница 46: ...Chapter 46 ...
Страница 47: ...47 5 Principles of Operation Analog Input Features 49 Digital I O Features 54 ...
Страница 48: ...Chapter 5 48 Figure 13 shows a block diagram of the DT9824 module Figure 13 Block Diagram of the DT9824 Module ...
Страница 56: ...Chapter 5 56 ...
Страница 68: ...Chapter 6 68 ...
Страница 69: ...69 7 Calibration Overview 70 Using the Auto Calibration Procedure 71 Using the Manual Calibration Procedure 72 ...
Страница 74: ...Chapter 7 74 ...
Страница 75: ...75 8 Troubleshooting General Checklist 76 Technical Support 78 If Your Module Needs Factory Service 79 ...
Страница 80: ...Chapter 8 80 ...
Страница 92: ...Appendix A 92 ...
Страница 98: ...Appendix B 98 ...