Chapter 4
48
Testing Single-Value Analog Input
To verify that the module can read a single analog input value, do the following:
1.
Connect a voltage source, such as a function generator, to analog input channel 0
(single-ended mode) of a DT9816 Series module. Refer to
for an example of how
to connect a single-ended analog input.
2.
In the Quick DataAcq application, choose
Single Analog Input
from the
Acquisition
menu.
3.
Select the appropriate DT9816 Series module from the
Board
list box.
4.
In the
Channel
list box, select analog input channel 0.
5.
In the
Range
list box, select the range for the channel (±10 V).
6.
Select
Single Ended
.
7.
Click
Get
to acquire a single value from analog input channel 0.
The application displays the value on the screen in both text and graphical form.
Содержание DT9816 Series
Страница 1: ...DT9816 Series UM 21336 X User s Manual Title Page ...
Страница 4: ......
Страница 8: ...Contents 8 ...
Страница 12: ...About this Manual 12 ...
Страница 13: ...13 1 Overview Key Hardware Features 14 Supported Software 16 Getting Started Procedure 19 ...
Страница 20: ...Chapter 1 20 ...
Страница 21: ...Part 1 Getting Started ...
Страница 22: ......
Страница 44: ...Chapter 3 44 ...
Страница 54: ...Chapter 4 54 ...
Страница 55: ...Part 2 Using Your Module ...
Страница 56: ......
Страница 57: ...57 5 Principles of Operation Analog Input Features 59 Digital I O Features 64 Counter Timer Features 65 ...
Страница 81: ...81 7 Troubleshooting General Checklist 82 Technical Support 84 If Your Module Needs Factory Service 85 ...
Страница 86: ...Chapter 7 86 ...
Страница 102: ...Appendix B 102 ...
Страница 108: ...Index 108 ...