Calibrating the Analog Input Circuitry
You can choose to calibrate the analog input circuitry automatically
(described on this page) or manually (described on
);
auto-calibration is the easiest to use and is the recommended
calibration method.
Using the Auto-Calibration Procedure
To calibrate the analog input subsystem automatically, perform the
following step:
1.
Click
Go
in the Auto Calibration box.
The zero and full-scale ranges are automatically calibrated.
Note:
If you are not satisfied with the analog input calibration, you
can load the factory default settings stored in the EEPROM by
clicking
Restore
in the A/D Configuration Factory Settings box.
Once you have finished this procedure, the analog input circuitry is
calibrated. If you are using a DT9805 or DT9806 function module, it is
recommended that you calibrate the thermocouple circuitry using the
instructions on
. Otherwise, you can calibrate the analog
output circuitry, if you wish, following the instructions on
Содержание DT9800 Series
Страница 1: ...DT9800 Series UM 17473 L User s Manual ...
Страница 4: ......
Страница 8: ...Contents viii Appendix A Specifications 119 Appendix B Connector Pin Assignments 131 Index 139 ...
Страница 13: ...1 1 Overview Features 2 Supported Software 7 Accessories 9 ...
Страница 76: ...Chapter 2 64 ...
Страница 77: ...65 3 Supported Device Driver Capabilities ...
Страница 88: ...Chapter 3 76 ...
Страница 112: ...Chapter 4 100 ...
Страница 113: ...101 5 Calibration Running the Calibration Utility 103 Calibrating the Analog Input Subsystem 104 ...
Страница 123: ...111 6 Troubleshooting General Checklist 112 Service and Support 115 If Your Board Needs Factory Service 118 ...
Страница 131: ...119 A Specifications ...
Страница 143: ...131 B Connector Pin Assignments ...
Страница 162: ...Index 150 ...