![Data Translation DT9800 Series Скачать руководство пользователя страница 121](http://html.mh-extra.com/html/data-translation/dt9800-series/dt9800-series_user-manual_511240121.webp)
Calibration
109
5
5
5
5
5
5
5
5
5
Note:
If you are not satisfied with the thermocouple calibration,
you can load the factory default settings stored in the EEPROM by
clicking
Restore
in the Thermocouple Calibration Factory Settings
box.
Once you have finished this procedure, the thermocouple circuitry is
calibrated. You can now calibrate the analog output circuitry, if you
wish, following the instructions in the next section.
Calibrating the Analog Output Subsystem
To calibrate the analog output circuitry of the DT9802, DT9802-EC,
DT9802-EC-I, DT9804, DT9804-EC, DT9804-EC-I, or DT9806 modules,
use an external precision meter available from vendors such as Fluke.
To calibrate the analog output circuitry, perform the following steps:
1.
Connect Analog Output 0+ (TB19) to the positive side of the
precision voltage meter.
2.
Connect Analog Output 0 Return (TB20) to the negative side of
the precision voltage meter.
3.
In the DAC0 Voltage box, click
0.000 V
.
4.
In the DAC0 Adjustment box, click the increment or decrement
arrows until your external meter display reads 0 V (within
0.005 V).
5.
In the DAC0 Voltage box, click
+9.375 V
and verify that your
external meter display reads +9.375 V (within ±12 mV.)
6.
Connect Analog Output 1+ (TB21) to the positive side of the
precision voltage meter.
7.
Connect Analog Output 1 Return (TB22) to the negative side of
the precision voltage meter.
Содержание DT9800 Series
Страница 1: ...DT9800 Series UM 17473 L User s Manual ...
Страница 4: ......
Страница 8: ...Contents viii Appendix A Specifications 119 Appendix B Connector Pin Assignments 131 Index 139 ...
Страница 13: ...1 1 Overview Features 2 Supported Software 7 Accessories 9 ...
Страница 76: ...Chapter 2 64 ...
Страница 77: ...65 3 Supported Device Driver Capabilities ...
Страница 88: ...Chapter 3 76 ...
Страница 112: ...Chapter 4 100 ...
Страница 113: ...101 5 Calibration Running the Calibration Utility 103 Calibrating the Analog Input Subsystem 104 ...
Страница 123: ...111 6 Troubleshooting General Checklist 112 Service and Support 115 If Your Board Needs Factory Service 118 ...
Страница 131: ...119 A Specifications ...
Страница 143: ...131 B Connector Pin Assignments ...
Страница 162: ...Index 150 ...