Calibration
121
Note:
Calibrating the PGH Zero setting in software is no longer required; the hardware
calibrates this setting automatically.
If you are not satisfied with the analog input calibration, you can load the factory default
settings stored in the EEPROM by clicking
Restore
in the Factory Settings box.
Using the Manual Calibration Procedure
To calibrate the analog input circuitry manually, do the following:
1.
From the main menu of the DT300 Series Calibration Utility, click
Configure
, and then
Board
.
2.
Select the name of the DT300 Series board to configure from the combo box, and then click
OK
.
3.
From the main menu of the DT300 Series Calibration Utility, click
Calibrate.
4.
Click
A/D
.
5.
In the Range box, select
Bipolar
, and then
Zero
.
6.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 0 V (within 0.0001 V for the DT321 and DT322 boards and within 0.0010 V for the
remaining DT300 Series boards).
7.
In the Range box, select
Bipolar
, and then
FS
(for full-scale).
8.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads +9.3750 V with the external reference (within 0.0001 V for the DT321 and DT322
boards and within 0.0010 V for the remaining DT300 Series boards).
9.
In the Range box, select
Unipolar
, and then
Zero
.
10.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 0 V (within 0.0001 V for the DT321 and DT322 boards and within 0.0010 V for the
remaining DT300 Series boards).
11.
In the Range box, select
Unipolar
, and then
FS
(for full-scale).
12.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads +9.3750 V with the external reference (within 0.0001 V for the DT321 and DT322
boards and within 0.0010 V for the remaining DT300 Series boards).
13.
Click
Quit
when you are finished calibrating the analog input circuitry.
Once you have finished this procedure, continue with
Содержание DT300 Series
Страница 1: ...DT300 Series UM 16501 T User s Manual Title Page ...
Страница 4: ......
Страница 10: ...Contents 10 ...
Страница 14: ...About this Manual 14 ...
Страница 15: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 20 ...
Страница 21: ...Part 1 Getting Started ...
Страница 22: ......
Страница 58: ...Chapter 4 58 ...
Страница 69: ...Part 2 Using Your Board ...
Страница 70: ......
Страница 106: ...Chapter 6 106 ...
Страница 124: ...Chapter 8 124 ...
Страница 125: ...125 9 Troubleshooting General Checklist 126 Technical Support 128 If Your Board Needs Factory Service 129 ...
Страница 130: ...Chapter 9 130 ...
Страница 156: ...Appendix C 156 ...
Страница 164: ...Index 164 ...