Calibration
123
Calibrating the Analog Output Subsystem
To calibrate the analog output circuitry, use an external precision meter. The following
sections describe how to configure for calibration using the supported meters and how to
calibrate the analog output circuitry.
Configuring for an External Meter
To calibrate DAC0 using an external voltage meter, do the following:
1.
Connect Analog Out 0+ (TB19) to the positive side of the precision voltage meter.
2.
Connect Analog Out 0 Return (TB20) to the negative side of the precision voltage meter.
To calibrate DAC1 using an external voltage meter, do the following:
1.
Connect Analog Out 1+ (TB23) to the positive side of the precision voltage meter.
2.
Connect Analog Out 1 Return (TB22) to the negative side of the precision voltage meter.
Continue with the next section.
Using the Calibration Procedure
To calibrate the analog output circuitry, do the following:
1.
From the main menu of the DT300 Series Calibration Utility, click
Configure
, and then
Board
.
2.
Select the name of the DT300 Series board to configure from the combo box, and then click
OK
.
3.
From the main menu of the DT300 Series Calibration Utility, click
Calibrate.
4.
Click
D/A
.
5.
In the D/A box, select
DAC0
.
6.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 0 V (within 0.0005 V).
7.
Select
+FS
and verify that the display reads +9.375 V (within ±12 mV for the DT322 and
within ±20 mV for the DT302 and DT304 boards.)
8.
In the D/A box, select
DAC1
.
9.
Click the increment or decrement arrows in the Manual Adjustment box until the display
reads 0 V (within 0.0005 V).
10.
Select
+FS
and verify that the display reads +9.375 V (within ±12 mV for the DT322 and
within ±20 mV for the DT302 and DT304 boards.)
11.
Click
Quit
when you are finished calibrating the analog output circuitry.
Once you have finished this procedure, the analog output circuitry is calibrated. To close the
Calibration Utility, click the close box in the upper, right corner of the window.
Содержание DT300 Series
Страница 1: ...DT300 Series UM 16501 T User s Manual Title Page ...
Страница 4: ......
Страница 10: ...Contents 10 ...
Страница 14: ...About this Manual 14 ...
Страница 15: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 20 ...
Страница 21: ...Part 1 Getting Started ...
Страница 22: ......
Страница 58: ...Chapter 4 58 ...
Страница 69: ...Part 2 Using Your Board ...
Страница 70: ......
Страница 106: ...Chapter 6 106 ...
Страница 124: ...Chapter 8 124 ...
Страница 125: ...125 9 Troubleshooting General Checklist 126 Technical Support 128 If Your Board Needs Factory Service 129 ...
Страница 130: ...Chapter 9 130 ...
Страница 156: ...Appendix C 156 ...
Страница 164: ...Index 164 ...