Insufficient Vacuum
If you notice insufficient vacuum on the probe
tips (e.g., devices are dropped), perform the Vac-
uum Check: see
some of the checks fail (red Xs on report when
the
details
check box is checked) then check that:
•
all hose fittings are tight and hoses are not
pinched or clogged.
Perform the probe blowout procedure; see
•
the air filters are not dirty. Replace air filters
if necessary. See
•
the input air to the FLX is at
.48-.55 MPa
(70-80 PSI).
If you can’t find the source of the problem, save
the detailed
Check Vacuum
report and send it to
Failures Picking and Placing
If probes drop devices or fail to pick and place
devices accurately, check the following:
•
check for insufficient vacuum; see previous
heading.
•
check the Z-Offset; see
•
check the module alignment; see
Содержание FLX500
Страница 4: ... Data I O FLX500 ii ...
Страница 12: ... Data I O FLX500 Chptr 1 8 blank page ...
Страница 28: ... Data I O FLX500 Chptr 2 24 blank page ...
Страница 40: ... Data I O FLX500 Chptr 4 36 blank page ...
Страница 42: ... Data I O FLX500 Chptr 5 38 Daily Sign off Clean the sockets page 10 Date Initials Notes ...
Страница 43: ...5 Sign off Charts Daily Sign off 39 Daily Sign off Date Initials Notes ...
Страница 44: ... Data I O FLX500 Chptr 5 40 Daily Sign off Date Initials Notes ...
Страница 45: ...5 Sign off Charts Daily Sign off 41 Daily Sign off Date Initials Notes ...
Страница 46: ... Data I O FLX500 Chptr 5 42 Daily Sign off Date Initials Notes ...
Страница 48: ... Data I O FLX500 Chptr 5 44 Weekly Sign off Date Initials Notes ...
Страница 52: ... Data I O FLX500 Chptr 5 48 blank page ...
Страница 57: ... FLX500 ...
Страница 58: ... FLX500 FLX programmable media experts ...