EASY
SENSE Q3+ and Q5+
DO185 (2)
•
Use the blue (
▲▼
) buttons to scroll the menu until the pointer is pointing at
System. Press the green
►
button to select.
•
Use the blue (
▲▼
) buttons to scroll the menu until
the pointer is pointing to the appropriate system
menu item. Press the green
►
button to select.
No Bluetooth
►
Sensor ON/OFF
Battery life
Sensor ON/OFF
The built-in sensors can be switched On or Off the display e.g. Sound and Temperature
switched off so only Light measurements are displayed.
►
Sound Off
Light On
Temperature Off
Use the blue (
▲▼
) buttons to scroll the menu until the
pointer is pointing at the appropriate Sensor. Press the
green
►
button to alter the selection from On to Off or
visa versa.
Press the red
button to return to the System menu.
If a hard reset is performed (page 39) it will automatically return all sensors to On.
Battery life
Battery life:
Awake: 13.5 hours
Easy Log: 42 days
Level: 93%
Note:
If
EASY
SENSE Q has been on charge, disconnect the power
supply and/or USB cable for at least 15 seconds before checking
values.
EASY
SENSE Q
will determine the power requirements of the connected sensors and
compare this with the charge level of the batteries. Bluetooth communication will
increase power consumption, so if you are using Bluetooth make sure it is switched on
before checking battery life.
The LCD will display an estimate of:
1.
Battery life Awake – the number of hours
EASY
SENSE Q could continuously
operate without a power supply attached.
2.
Easy Log – if
EASY
SENSE Q is set to record in Easy Log mode and is operating on
battery power, it will auto switch off between samples to reduce battery consumption.
Battery life Easy Log indicates the theoretical number of days that
EASY
SENSE Q
could record in this mode.
3.
The percentage capacity charge level of the battery.
The LCD display will show ‘Battery is charging’ if power is being supplied to
EASY
SENSE Q
(via a power supply or USB connection).
Press the red
button to return to the System menu.
Note:
Ideally
EASY
SENSE Q should be stored with at least 40% or more charge. Do not store
EASY
SENSE Q with its battery fully discharged or with sensors connected. Some reserve charge is needed
to retain data and settings in memory. See page 26 for powering information.
©