background image

  

 

 

 

Stay still while recording data. Movement e.g. raising and lowering a hand, will change blood 
flow in the finger and give erratic readings. 
  
 

If the finger/earlobe clip ‘loses contact’ 
with the pulse, the reading may 
momentarily drop to zero. 
 

 
 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 

The light sensitive detector (LSD) in the clip can be sensitive to high levels of ambient infrared 
light e.g. bright sun or room lights. If the light interferes with the detector the Sensor may 

‘lose 

contact

’. Stop logging, and reduce the light level e.g. by covering the sensor clip before starting 

to log again.  

Note:

 If you alter the light level while logging the sensor will need to adjust to the new conditions. 

 

Avoid applying pressure to the sensor clip e.g. by resting the finger clip on a work surface.  

It is possible to alter heart rate by changing breathing. Encourage the test subject to breath 
normally. 

This Sensor is 

not

 intended for medical diagnosis. Do not be alarmed if results do not 

correspond to quote numbers, these represent typical averages. It is quite normal for the 
heart to occasionally miss a beat. 

 

It is possible for the dicrotic notch 
to be recorded as a separate pulse, 
and so give higher readings than 
expected. The effects are usually 
more noticeable if the clip is 
attached to the finger; try using with 
the clip attached to the ear lobe.  

 

 

 

 

This sensor is not waterproof. Clean the clip by wiping with a cloth that has been rinsed in 
disinfectant.  

 

 

 

Contact lost  

Communication 
re-established 

 

 

Example of the depth of dicrotic notch 

that may affect Heart rate data. 

Dicrotic notch 

 