______________________________________________________
______________________________________________________ MODEL
MODEL 2500
2500
3.
The following illustrated message of the MODEL 2500 LCD screen indicates
that a PROM has been changed since installation, or is being read
improperly. The message on line 2 (XXXX) is the location of the error.
CHECKSUM ERROR
XXXX
4.
If the MODEL 2500 is not configured for an application:
-
Alarm contact closure inside the case can be heard opening and
closing.
-
Red LED light on the front panel flashes on and off.
-
Following illustrated message of the MODEL 2500 LCD screen is
displayed.
AWAITING CONFIG
PX BXXXX IDXXX
Line 2 of the display message indicates:
-
PX is the Port (for example, P2 or Port 2).
-
BXXXX is the baud rate (for example, B2400 or 2400 baud).
-
IDXXX is the COMMID (for example, ID001 or COMMID
001).
If this message is displayed when power is applied to the MODEL
2500, the instrument must be configured before proceeding.