Commissioning
6.3 Start-up routine
SONO 3500 CT
Operating Instructions, 07/2019, VI.SH.P2.02
39
Menu
Parameter
Display example
Comments
Menu 5
Display test
Check of all segments. Dis-
play toggles between all
segments on/off.
6.3
Start-up routine
1. Power-up device.
2. Reset battery status indicator as described below.
Resetting the battery counter
Note
The reset can also be carried out via SIMATIC PDM.
When a new battery is installed and the plug is connected, the transmitter start-up routine
begins. The display shows the active software version, e.g. 2.03.
After ten seconds the message "reset.bat" will appear.
1. Press push button within six seconds to reset the internal battery counter. The message
"accept" will appear.
2. Press push button again within six seconds in order to reset the internal battery counter.
The battery indicator now shows full. If the push button is not pressed again, the battery
indicator will show "Low".
Figure 6-2
Reset internal battery counter
* Firmware version, here version 2.03
Содержание SONO 3500 CT
Страница 1: ...Ultrasonic Flowmeter Operating Instructions SONO 3500 CT 07 2019 Edition ...
Страница 2: ......
Страница 8: ...Table of contents 6 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 14: ...Introduction 1 5 Further Information 12 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 38: ...Electrical connection 5 5 Sealing of SONO 3500 CT 36 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 50: ...Commissioning 6 4 Commissioning via PDM 48 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 54: ...Functions 7 4 Hardware key 52 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...
Страница 64: ...Service and maintenance 8 7 Battery disposal 62 SONO 3500 CT Operating Instructions 07 2019 VI SH P2 02 ...