Input data

Battery charge

Nominal input voltage

24 V DC

Status display                                       Red/Green  LED Bargraph

Input voltage range

18 V DC – 30 V DC

Switching output                                           Relay 

(Preset: switching output - active high)

30 V AC/DC , ≤ 100 mA

Current consumption

14 A

Output data (mains operation)

Signal supply

Nominal output voltage

24 V DC

Switching voltage                                   Buffered ≤ 24 V DC 200 mA

Output voltage range (depending on the 

input voltage)

18 V DC – 30 V DC

Switching current                Short-circuit-proof 

Output current with convection cooling

Remote control

Nominal output current                   Permanent

Power Boost

SFB technology                                              12 ms

10 A

Version 1

Output R1 to input R2                Plug-in-bridge 

15 A
60 A

Version 2

Input R2

Input current R2

24 V DC

200 mA

Output data (battery operation)

Nominal output voltage

24 V DC

Output voltage range (depending on the 

battery voltage)

19,2 V DC – 27,6 V DC 

(UOUT = UBAT - 0,5 V DC)

Energy storage

Output current with convection cooling

Battery type

Panasonic UP-VW1220P1

Memory medium     Lead rechargeable 

battery module, 3.4 Ah    
Nominal voltage

24 V DC

End-of-charge voltage (temperature 

compensated)

24 V DC – 29 V DC

Capacitance

3,4 Ah

Charging current ICHARGE

2,88 A

Battery service life

6 – 9 (20 °C)

Latest startup

Months 6 (0 °C – 20 °C)
Months 6 – 3 (20 °C – 30 °C)
Months 3 – 1 (30 °C – 40 °C)

General data

Degree of protection / Protection class

IP20 / III

Degree of pollution

2

Ambient temperature (operation)

0 °C – 40 °C

Ambient temperature (storage/transport)

-15 °C – 40 °C

Humidity at 25 °C, non-condensing

≤ 95 %

Dimensions (W/H/D)

120 x 169 x 125 mm

Weight

3,8 kg

Technical Data

Output current with convection cooling

Nominal output current                   Permanent

Power Boost

SFB technology                                              15 ms 

10 A

15 A

65 A

Buffer time (IOUT =      )

180 min. (1 A) / 10 min. (10 A)

Connection types

Connection in parallel                                      Yes 2
Connection in series                                          No 

Signaling

Power In OK

Status display                                                 Green LED

Alarm

Status display                                                      Red 

Switching output                                           Relay

(Preset: switching output - active low)

LED

30 V AC/DC , ≤ 100 mA

Battery mode

Status display                                                Yellow

Switching output                                           Relay

(Preset: switching output - active high)

LED

30 V AC/DC , ≤ 100 mA

IN

IN

IN