manualshive.com logo in svg
background image

3.1  Calibration steps

3.2  Zero calibration

Prior to calibration, the sensors must be supplied with power voltage without interruption for run-in 

and stabilization. 
The run-in time depends on the sensor element and is shown in the following tables, as well as the 

other relevant information:

First enter in Service Mode 
1.  Press Enter to enter in the menu and press down arrow until the Installation & Calibration menu
2.  Press Enter and Service Mode OFF is shown
3.  Press Enter, enter the password ****, press Enter and down arrow to change the status from OFF to 

ON and then press Enter again.

  When the unit is in Service Mode the display yellow LED is blinking.

From the Installation & Service menu, by using the down arrow scroll until the Calibration menu and 

press Enter.

The type of gas sensor is displayed.

By using the Enter and up/down arrow keys set the calibration gas concentration in ppm:
•  for CO2 sensor, select 10000 ppm which corresponds to 50% of the sensor measuring range
•  for HFC sensor, select 1000 ppm which corresponds to 50% of the sensor measuring range
•  for PE sensor, select 250 ppm which corresponds to 50% of the sensor measuring range

Select the Zero calibration menu.
In case of CO2 sensor, the Zero Calibration has to be executed by exposing the sensor to pure 

Nitrogen, same gas flow.
Before executing the zero calibration the specified warm-up times must be strictly observed before 

starting the process.
Connect the calibration gas cylinder to the sensor head by using the calibration adapter 148H6232.

Open the calibration gas cylinder flow regulator. 
During calculation an underscore in line two, runs from left to right and the current value drops to zero.

When the current value is stable press Enter for saving the calculation of the new value.

“SAVE“ is displayed, as long as the function is executed.

After the value has been successfully stored, a square appears on the right for a short time = zero 

point calibration is finished and new zero offset has been stored with success.
The display automatically goes to the display of the current value.

Sensor Element

Gas

Run-in time 

calibration (h)

Warm-up

time (s)

Flow rate

(ml/min)

Gas application 

time (s)

Infrared

Carbon dioxine

1

30

150

180

Semiconductor

HFC

24

300

150

180

Pellistore

Combustible

24

300

150

120

Fig. 3

Operating Guide | DGS functional tests and calibration procedure

© Danfoss | Climate Solutions | 2022.10

AQ430229665343en-000101 | 5

Содержание DGS-IR CO2

Страница 1: ...Operating Guide DGS Functional tests and calibration procedure ...

Страница 2: ...he unit must be carried out by a suitably qualified technician and must be done in accordance with this guide in compliance with locally applicable guidelines and regulations Recalibration and part replacement in the field maybe implemented by a qualified technician and with the appropriate tools Alternatively the easily removable sensor element may be replaced There are two concepts that need to ...

Страница 3: ...or calibration or accuracy checks of the sensor These require actual gas calibration not bump testing with ampoules For bump tested expose the sensors to test gas R134A CO2 etc The gas should put the system into alarm The purpose of this check is to confirm that gas can get to the sensor s and that all the alarms present are functional For bump test can be used Gas Cylinder or Gas Ampoules see Fig...

Страница 4: ...flow controller Calibration adapter with tube code 148H6232 Note about test gas bottle for calibration see Fig 1 because the transportation of gas ampoules and cylinders gas is regulated by many governments throughout the world is suggested to source them from local dealers Before to perform the calibration connect the Handheld Service Tool 080Z2820 to the DGS device 1 Danfoss 148H153 JP5 5 4 2 1 ...

Страница 5: ...responds to 50 of the sensor measuring range Select the Zero calibration menu In case of CO2 sensor the Zero Calibration has to be executed by exposing the sensor to pure Nitrogen same gas flow Before executing the zero calibration the specified warm up times must be strictly observed before starting the process Connect the calibration gas cylinder to the sensor head by using the calibration adapt...

Страница 6: ...ce value of the set calibration gas concentration press Enter for finishing the calculation of the new value After the value has been successfully stored a square appears on the right for a short time Gain calibration is finished a new gain offset has been stored with success The display automatically goes to the display of the current ppm value During the calculation phase the following messages ...

Страница 7: ...y Calibration tool no This report was created automatically and is also valid without a signature Calibration compiles with ISO IEC 17025 Date Alternative Cal Gas Employee ppm C mbar ml min ppm ppm h Alt part no Humidity Actual calibration value Days left until maintenance Address Company Manufact no Operating Guide DGS functional tests and calibration procedure Danfoss Climate Solutions 2022 10 A...

Страница 8: ... Danfoss Climate Solutions 2022 10 AQ430229665343en 000101 8 ADAP KOOL ...

Отзывы: