transmitted to the wire-speed of 1 Gigabit per second. Therefore, inter-packet delay is typically only used when 
multiple detectors are connected to the same port of the network card, through an Ethernet switch. 
It is important to consider that inter-packet delay inserts a minimum delay between image packets to spread packet 
transmission over a longer period of time. This can directly impact system latency as more time than could be 
necessary is put in between those packets. The best approach for real-time imaging is to dedicate a different network 
port to each detector. This way, the inter-packet delay can be eliminated in many cases. 
Some network equipment also supports the optional IEEE802.3 PAUSE mechanism. This is a low-level handshake 
to ensure the receiver of the packets is not overwhelmed by the amount of data. It can propagate a pause signal back 
to the transmitter, asking to momentarily stop the data transmission (with a possible impact on the overall system 
latency). Again, by combining network equipment that can operate at wire-speed and allocating a different network 
interface port for each detector in the system, we can ensure these pause requests will not be used. 
 
 

Connecting  

1. After the install is complete, plug in the detector communication cable.  

2. Power up the detector.  

3. After powering up the detector, the application automatically detects the detector. This may take up to 1 minute. 
Wait until the detector connected indicator in the task bar (shown below) has confirmed that a connection has been 
established (the red cross over the camera icon disappears).  
 

 

 

Checking detector status and information in Gigabit Ethernet  

When you add or remove a detector from the network, it is automatically detected by the GigE server. A manual 
network scan can also be launched by right-clicking the camera icon and selecting “Scan Network” from the menu 
options. This action will refresh the information of the device connected to the network. 
 
After the detector connection has been established, the status of the connection can be displayed by double-clicking 
the camera icon in the task bar. The device status window displays the serial number of the detector that has been 
detected on the network and its current status. Additional information with respect to the detector model, serial 
number and MAC address is also displayed.  
 
The user can also view the detector and the PC IP addresses in this window. The default assigned detector IP address 
can be modified using the Teledyne DALSA Network Configuration Tool, which is part of the installation package. 
 

 

 

Содержание Shad-o-Box 1K HS

Страница 1: ...Teledyne DALSA Shad o Box 1K 2k HS Detector User Manual December 11 2018 www TeledyneDALSA com ...

Страница 2: ...uit and electronics technology software and highly engineered semiconductor wafer processing Products and services include image sensor components electronic digital cameras X ray detectors vision processors image processing software and semiconductor wafer foundry services for use in MEMS high voltage semiconductors image sensors and mixed signal CMOS circuits Contact Information Please feel free...

Страница 3: ...Body and Front Cover 7 Periodic Maintenance 7 Calibration 7 5 Specifications 8 6 Connectors Cables and Status Indicators 9 7 Setting up the Detector 10 GigE Network Adapter 10 Connect Ethernet cable 10 Connect power cable 10 Status Light 10 Software Installation 10 8 Software Installation and Setup 11 Operating Systems Support 11 Obtain software 11 Sapera LT and CamExpert Installation 11 Optimizin...

Страница 4: ...Shad o Box 1K 2K HS User Manual p 4 Shuttering 17 11 Mechanical Drawings 18 Revision History 20 ...

Страница 5: ...DALSA assumes no responsibility for the proper installation of X ray shielding X ray shield enclosure testing or safe and prudent operation of the camera system in the end user s installation It is the end users responsibility to ensure that local and federal guidelines regarding the installation and operation of X ray sources are followed For those parties supplying the power requirements to the ...

Страница 6: ...old by Teledyne DALSA as replacement parts may result in increased emission or decreased immunity of the Shad o Box HS Detector Static precautions Observe proper ESD static control procedures when handling system components The use of properly grounded wrist strap is highly recommended Installation Precautions Never connect or disconnect cables while power to the detector is on Damage to the detec...

Страница 7: ...etector body surface Wipe off with a clean soft cloth Do not use any harsh cleansers or solvents that may damage the paint or other finishes on the detector In the event of detector exposure to bio hazardous materials proper cleaning procedures should be undertaken prior to removal or maintenance of the detector Periodic Maintenance No periodic maintenance or service is needed Calibration To have ...

Страница 8: ... 1 Non linearity 10 90 FS 1 5 Readout period 3 45 ms Output data rate 40 MHz 1 dark current doubles approx every 8 C 2 ADU Analog Digital Unit 1 LSB Least Significant Bit 3 time required to transfer image from sensor to camera memory Camera Specifications Units Typical supply voltage 12 0 Volts Supply voltage range 11 to 13 Volts Maximum supply current 1 0 Amps Typical power dissipation 10 Watts D...

Страница 9: ...type FHG 0B 302 CLAD52Z straight type FGG 0B 302 CLAD52 Detector Status LED The detector is equipped with a color LED used to display the status of the detector s operation The table below summarizes the operating states of the detector and the corresponding LED states Color of Status LED Meaning 1 Off No power 2 Green Detector is operational pin Description 1 Supply voltage 12V 2A 2 Ground The po...

Страница 10: ... Intel PRO 1000 MT adapter Review the NIC documentation concerning any special driver required for Windows Install the PCI bus Gigabit NIC as described by the NIC manufacture s documentation Connect Ethernet cable Connect the Ethernet cable from the detector to the computer Ethernet jack Secure the RJ 45 connector to the detector Connect power cable Connect a power cable from the detector to a 12V...

Страница 11: ...ion Sapera LT and CamExpert Installation A link to download the software will be emailed to you after submitting the application Complete the download process and run the installation file e g SaperaLT80xSDKSetup exe CamExpert is an image viewer designed to evaluate your detector The Sapera SDK is a set of development tools and documentation for creating your own software for interfacing to the pr...

Страница 12: ...arger than 9000 bytes For this reason we advise to set the packet size to 9014 bytes jumbo packet enabled Interrupt Moderation Normally each time a packet is received by the network card the associated driver will receive an interrupt Obviously when the packet rate is very high that is at high transfer rate which is common for GigE Vision systems this represents significant overhead Most network c...

Страница 13: ...n conditions the host PC system CPU may be busy with tasks other than the imaging application Incoming image packets remain in the PC memory allocated to store packets instead of immediately being copied into the image buffer By increasing the number of NIC network interface card receive buffers more incoming image packets can be stored by the NIC before it must start discarding them This provides...

Страница 14: ...e used to manage flow control i e the speed at which stream packet can be output to the network This is useful when connecting multiple detectors to the same port of the network card or when the network card Ethernet switch if used is simply too slow to process those packets A careful selection of equipment will ensure that the network equipment is fast enough to handle data ...

Страница 15: ...requests will not be used Connecting 1 After the install is complete plug in the detector communication cable 2 Power up the detector 3 After powering up the detector the application automatically detects the detector This may take up to 1 minute Wait until the detector connected indicator in the task bar shown below has confirmed that a connection has been established the red cross over the camer...

Страница 16: ...the grabbing process generates multiple frames by operating in a continuous mode To snap an image Click the Snap button The Detector has now snapped a single frame The image appears in the GUI window To grab multiple images Click the Grab button The detector is now armed and ready to acquire images in continuous mode To stop a Grab process the user must press the Freeze button After the Freeze but...

Страница 17: ...e interval is marked with the dark green rectangle Row Pixel Array Time ReadOut Integration Time Frame N Frame N 1 Trigger Input Trigger Output The period of time in between successive frame read outs dark green is used to fire the pulsed X ray source The width of this time period can be programmed via the detectors command interface by adjusting the ExtendedExposure us parameter The ExtendedExpos...

Страница 18: ...Shad o Box 1K 2K HS User Manual p 18 11 Mechanical Drawings Shad o Box 1K HS Model all measurements are in mm ...

Страница 19: ...Shad o Box 1K 2K HS User Manual p 19 Shad o Box 2K HS Model all measurements are in mm ...

Страница 20: ...Box 1K 2K HS User Manual p 20 Appendix A Revision History Revision Number Change Description Revision Date 01 Draft release May 10 2016 02 New address January 12 2018 03 Added SB2kHS detector December 11 2018 ...

Отзывы: