Linea HS 32k CMOS TDI Cameras
Camera Performance and Features
27
Direction Change Time
The direction change time between forward and reverse is < 1 ms.
Setting the correct scan direction
Whether the scan direction is set correctly can easily be seen in live imaging. An image will appear
“normal”, sharp and focused. If the optical setup is not properly focused, blur will occur in both,
horizontal (cross-scan) and vertical (in-scan), directions.
If blur occurs only in scan direction (see below), the scan direction is set incorrectly.
Figure 10. Image with incorrect scan direction