2.2.  Flammable gases detector (DEX CAN transmitter) 

 
To perform a correct calibration of a flammable gases detector you need : 

 

1.  A calibration gas cylinder (concentration between 1/3 and 2/3 of measuring range); 

2.  A pressure reducer with flowmeter; 

3.  An injection nozzle; 

4.  A Teflon or polyurethane flexible hose (Ø 7mm); 

5. A 

voltmeter; 

6.  A measurement plug (pitch = 2.54mm); 

7.  An insulated flat screwdriver (width 3.5mm). 

 

To get the injection nozzle or the measurement plug please contact Dalemans. 
 
For applying gas to the detector, the flexible hose has to be plugged at one end to the cylinder flowmeter, and at 
the other end to the injection nozzle. The injection nozzle must be fitted in the detection head opening (

).  

Figure 8

Figure 8 : Applying  gas to the detector

 

Man-CanDetectors-Cal-En-V10.doc

 

Can Bus Gas Detectors - 
C lib

ti

 

 

Figure 6 : Calibration gas 

Figure 7 : Measurement plug 

 

Cylinder 

Flowmeter 

Hose 

Hose 

Injection 

nozzle 

Detection 

head 

Pitch 2.54mm

 

e

 

Depending on the type of installation, the DEX CAN transmitter is located in the detector junction box (Figure 
9) or inside the OctoBUS 64 central unit (Figure 10). 

 

 

Figure 9 : DEX CAN detector 

Figure 10 : DEX CAN transmitter with remote detector 

OctoBUS 64 central unit 

 

Remote detector

DEX CAN Transmitter 

Detection head 

DEX CAN 

Transmitter 

 
To perform calibration always proceed in the following order : 

 

1.  Set the operating voltage of the cell; 

2.  Set the zero of DEX CAN (Zeroing); 

3.  Adjust sensibility of DEX CAN (Gain). 

Содержание octoPRO

Страница 1: ...CAN Bus System Calibration manual for Toxic Flammable gases detectors Gas detection w w w d a l e m a n s d e t e c t i o n c o m...

Страница 2: ...mans S A can not be liable for direct or indirect damages arising out of the non observance of its instructions Use only Dalemans original parts when you assume the maintenance of the material as desc...

Страница 3: ...ON 2 2 1 Toxic gases oxygen detector DTX CAN 2 Zeroing the DTX CAN 3 Adjusting gain of DTX CAN 3 2 2 Flammable gases detector DEX CAN transmitter 4 Set the cell operating voltage 5 Zeroing the DEX CAN...

Страница 4: ...19 54 55 34 Email info dalemans detection com In a Dalemans CAN bus system based on OctoBUS 64 central measuring unit to perform the complete calibration procedure of the detectors you need the hand h...

Страница 5: ...rewdriver width 1mm In order to avoid ambient air or air currents from corrupting the calibration atmosphere around the cell always use the suitable injection cap Just place it instead of the metal de...

Страница 6: ...Instruction Manual for measure function Figure 5 Adjusting GAIN on DTX CAN Adjusting gain of DTX CAN Detector must have been zeroed as described here above On the cylinder set flow rate between 0 4 t...

Страница 7: ...ion nozzle The injection nozzle must be fitted in the detection head opening Figure 8 Figure 8 Applying gas to the detector Man CanDetectors Cal En V10 doc Can Bus Gas Detectors C lib ti Figure 6 Cali...

Страница 8: ...CAN transmitter Power on the DEX CAN transmitter and wait for at least 15 sec Set the voltmeter on Vdc position and connect it between terminals A and P of the DEX CAN transmitter Note the transmitter...

Страница 9: ...ction Manual for measure function Adjusting gain of DEX CAN Detector must have been zeroed as described here above Visualize the detector measure on the OctoPRO display press F2 or on the OctoBUS 64 c...

Страница 10: ...NOTES Gas detection...

Страница 11: ...il internet R u e J u l e s M l o t t e 2 7 4 3 5 0 R e m i c o u r t B e l g i u m Tel 3 2 0 1 9 5 4 5 2 3 6 Fax 3 2 0 1 9 5 4 5 5 3 4 Gas detection info dalemans detection com www dalemans detection...

Отзывы: