ZX-5
Ultrasonic Thickness Gauge
35
7.8 Zero
There are two transducer zeroing options available in the
ZX-5
; auto and manual.
The AUTO zero can be performed automatically without using a reference standard
to zero the gauge (off block), while the MANUAL option requires a reference standard
like the battery disk at the top of the gauge (on block). Additionally, the gauge can be
set to use another reference standard if needed. Refer to page 15 for a complete
explanation of the probe zero options.
The procedure to select the zero option only, is outlined below:
Zero (Auto/Manual)
1) Press the
key to access the menu items/features.
2) Use the
keys to scroll through the items/features until the
ZERO
feature is being displayed.
3) Press the
key to edit the zero setting. The edit icon will be illuminated
and flashing.
4) Use the
keys to select the auto/man option.
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes.
5) When the desired
ZERO
setting is displayed, press the
key to set the
status and return to the measurement screen.
7.9 Velocity (VX)
When the velocity setting (
VX
) is enabled, the
ZX-5
will display will the material
velocity as the primary measurement quantity instead of dimensional thickness. The