Dakota Ultrasonics
20
3) Press the
key to enter the calibration edit screen displaying the current
measurement value.
4) Use the
keys to scroll to the known thickness value.
Note:
The longer the keys are pressed and held, the faster the value will
increment/decrement.
Note:
Pressing the
key prior to pressing the
key will abort the cal
routine without saving any changes.
5) Once the known thickness value is being displayed, press the
key to
display “
1 of 2”
, which sets the zero value and returns to the measurement
screen.
Note:
The internal zero of the ZX-3 is now set. The procedure above can be
used to set the internal zero of the ZX-3 to use any reference standard as
the ‘probe zero’ standard if desired.
6) Repeat steps 2-4 on the second test point/location.
7) Press the
key to display the calculated velocity edit screen.
Note:
The calculated velocity can be edited, if needed, by pressing the
keys to scroll and edit the velocity value.
8) Press the
key to set the calculated material velocity and return to the
measurement screen.
Note:
CHECK YOUR CALIBRATION! Place the transducer back on the
calibration points. The thickness readings should now match the known
www.calcert.com