ZX-5
Ultrasonic Thickness Gauge
31
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes to the HI limit value.
9) When the target
HI
limit is displayed, press the
key to set the value
and return to the measurement screen.
7.4 Differential
The
Differential Mode
of the
ZX-5
provides the user with the ability to set a nominal
value, according to what the expected thickness should be, and measure the +/-
difference from the nominal value entered. This feature is typically used in QA
incoming inspections on pipes, plate stock, coils, etc.
The steps below outline how to enable and enter the nominal value to use this
feature:
Differential
1) Press the
key to access the menu items/features.
2) Use the
keys to scroll through the items/features until the
DIFF
feature is being displayed.
3) Press the
key to edit the differential status. The edit icon will be
illuminated and flashing.
4) Use the
keys to toggle differential on/off.
5) When the desired
DIFF
setting is displayed, press the
key to set the
status and edit the
NOMINAL
value.