PZX-7 DL
Ultrasonic Thickness Gauge
33
2) Use the
keys to scroll through the items/features until the
ALRM
feature is being displayed.
3) Press the
key to edit the alarm status. The edit icon will be illuminated
and flashing.
4) Use the
keys to toggle alarm on/off/beep.
5) When the desired
ALRM
status is displayed, press the
key to set the
status and edit the
LO
limit option.
6) Use the
keys to scroll the
LO
limit value to the target value.
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes to the LO limit value.
7) When the target
LO
limit is displayed, press the
key to set the value
and advance to setting
HI
limit option.
8) Use the
keys to scroll the
HI
limit value to the target value.
Note:
Pressing the
key prior to pressing the
key will abort to the
measurement screen without saving changes to the HI limit value.
9) When the target
HI
limit is displayed, press the
key to set the value
and return to the measurement screen.
8.4 Differential
The
Differential Mode
of the
PZX-7 DL
provides the user with the ability to set a
nominal value, according to what the expected thickness should be, and measure the
+/- difference from the nominal value entered. This feature is typically used in QA
incoming inspections on pipes, plate stock, coils, etc.
Содержание PZX-7DL