DAIHAN Scientific, Smart-Lab Temp./ Humidity Chamber 36/66
-
Limit of offset values are -10.0
℃
~ +10.0
℃
, and current temperature after setting offset
value displays the value that includes the offset.
-
Once you set the offset value, the value is stored and the offset value is reflected in the
current temperature even you press ON/OFF button.
3.3.6. Error Message
-
If happen a error, Error notice will be showed in the display and if you click it, you can see
what happened in detail.
* When malfunctions in the Touch Screen occur, please press the Reset Button by inserting an
appropriate tool (such as a paper clip) into the Reset hole, which is located in the right frame of
the Touch Screen as shown in the figure below. The Touch Screen is rerun by pressing the
Reset Button.
ERROR
Message
Causation
Solution
Temp. Error
When display show unreasonable
temp. value by out of order of
temp. sensor
Call the A/S Center.
Humi. Error
When display show unreasonable
Humi. value by out of order of
Humi. sensor
Call the A/S Center.
Out of
Control Error
When present temperature over 15
degree(can adjust this limit in the
User Config. manu)
Check the set value of it and refrigeration
system
Water Error
Notifies when there is insufficient
amount of solution in Water Tank.
Please until appropriate line fill in the
solution of inside Water Tank.
Door Open
When over 1 min. after open the
Door.
Check status of the Door and if still
happen it even door closed, call the A/S
center.
Revolution
Error
When the Communication error
problem of equipment and Touch
Board.
Please check the connection of the Touch
Board.