<Table 3.3.1 > ERROR Message Table
-
When entering into the compensation value setting mode during the Heating operation, the
Heating operation that is operating is stopped. After setting the compensation value, the
Heating operation is re-conducted in the Set Temperature Mode.
-
After setting offset value of one temperature range, Setting back the Offset value for correct
temperature control in case of the use another temperature range. Is users choice case for
accurate temperature control of test.
-
If the PUSH button of Jog Dial Knob is pressed and held for 2-3 seconds, the setting value is
saved and this saved value is maintained even when the power is turned on again.
-
If pressing MAIN button, it is switched to the Set Temperature Mode.
③
MAIN - Safety Locking
-
If pressing MAIN button in the Set Temperature Mode, it is switched to the Locking mode.
-
In the Locking Mode, the Turning by the Jog-Dial Knob or Push button do not applied, and LCD
Display and LED lamp indicate the status that is currently operating.
-
If pressing once the MAIN button, it is entered into the Set Operation Time Mode.
* Automatic Operation Function
-
In the use of the product, if the power supply has been interrupted and then is reapplied, it has
the function that protects the experimental results and samples by automatically reactivating as
the setting of final operation.
-
In Jog-Dial operation method, support this function by activating the setting temperature stored
through the saving method of the set points.
-
When reapplying the power supply according to the problem on safety or user’s intention, if you
want to disable the automatic operation function, set up the temperature to zero (0) according
to the instruction of Jog-Dial.
* ERROR Message
ERROR
Message
Cause
Remedy
Error 1
Informs the time to be displayed the temperature of
abnormal value due to the abnormality of
Temperature Sensor.
Please check the temperature of
the heated top plate.
< Fig 3.3.5. Safety Locking Mode >
Temp PV
: 27
℃
Time PV : 00:00
Present Temp
Remain Time