30
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-
SUPPORTING
Guidance and manufacturer’s declaration - electromagnetic immunity
The PM-150V is intended for use in the electromagnetic environment
specified below. The customer or the user of the PM-150V should assure
that it is used in such an environment.
IMMUNITY
test
IEC
Compliance
level
Electromagnetic environment - guidance
Radiated
RF
IEC
61000-
4-3
3 V/m
80
MHz
to 2.5
GHz
3 V/m
Portable and mobile RF communica-
tions equipment should be used no
closer to any part of the PM-150V,
including cables, than the recommen-
ded separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
800 MHz to 2,5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommen-
ded separation distance in meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an electro-
magnetic site survey,a should be less
than the compliance level in each
frequency range.b
Interference may occur in the vicinity
of equipment
marked with the following symbol:
P
d
3
.
2
=