Operating condition
Checking points
No electric power
-Plug correctly inserted into outlet?
-Correct the voltage of product with your local
voltage?
-Fuse short?
Power on but not-operating
-Pull out power plug and reinsert it into outlet.
Noisy during operating
-Horizontally installed?
-Any material on the device?
No air in the hose
-There any bending place?
-Hose plug correctly inserted into the device?
When the pressurization stays for
only 10 seconds at any chamber
and skips to the next chamber,
It is evidence that the chamber is leaking. The
chamber should be repaired but user can user can
continue to use regardless of the trouble chamber
until the cuff is repaired. (Pressure sensor perceives
the leaking automatically and operation program
makes keep for only 10 sec on the trouble chamber
and skip to the next chamber.