I
NSTRUCTION
M
ANUAL
3601 E. 34th St. Tucson, AZ 85713 USA Tel. +1
-
520
-
882
-
6598 Fax +1
-
520
-
882
-
6599 email: [email protected] Web: https://www.metallographic.com
Please read this instruction manual carefully and follow all installation, operating and safety guidelines.
GIGA
-
S Vibratory Polisher
8.0 Troubleshooting
22
Pry open fuse holder
with small flat head
screwdriver
Replace fuse
(10 amp fast blow)
More extensive troubleshooting, repair guides, videos, & parts list are
provided online at www.metallographic.com or
https://www.metallographic.com/PACE
-
service/GIGA
-
service.html
Problem
Cause
Solution
No power or function
a. Unit is disconnected
from main electrical power
supply
a. Verify electrical source
and connection
b. Main power switch is off
b. Turn on main power
switch
c. Blown fuse
c. Replace fuse
Samples won
’
t rotate
a.
Frequency isn
’
t properly
set
a. Reset frequency
according to procedure in
section 4.4
b. Voltage isn
’
t properly set
b. Increase voltage or
speed