66
12.2. GUIDANCE AND DECLARATION FOR ELECTROMAGNETIC IMMUNITY
PHENOMENON
BASIC EMC STANDARD
OR TEST METHOD
IMMUNITY TEST LEVEL COMPLIANCE LEVEL
Electrostatic discharge IEC 61000-4-2
± 8 KV contact
± 2 KV, ± 4 KV, ± 8 KV,
± 15 KV air
± 8 KV contact
± 2 KV, ± 4 KV, ± 8 KV,
± 15 KV air
EM fields of radiated RF
IEC 61000-4-3
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
3 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Fields in the vicinity
f r o m R F w i r e l e s s
c o m m u n i c a t i o n s
equipment
IEC 61000-4-3
See table
See table
Fast / saved electrical
transients
IEC 61000-4-4
alternating current
power input
± 2 kV
100 kHz repetition
frequency
± 2 kV
100 kHz repetition
frequency
IEC 61000-4-4
signal input/output
± 1 kV
100 kHz repetition
frequency
± 1 kV
100 kHz repetition
frequency
Outbreak Line by line
IEC 61000-4-5
± 0,5 kV, ± 1 kV
± 0,5 kV, ± 1 kV
Land line outbreak
IEC 61000-4-5
± 0,5 kV, ± 1 kV, ± 2 kV ± 0,5 kV, ± 1 kV, ± 2 kV
Magnetic fields at the
stated feed frequency
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
NOTE 1 At 80 MHz and 800MHz, the higher frequency range is applicable.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
NOTE 3 Conducted Disturbances Induced by RF Fields Test not applicable, as the equipment is
internally energized and cannot be used during battery charging.