MNL-00008 [5]
Proprietary and Confidential
Page
19
of
55
mW/cm
2
mJoules
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
7.0
240
01:36
01:20
01:09
01:00
00:53
00:48
00:44
00:40
00:34
280
01.52
01:33
01:20
01:10*
01:02
00:56
00:51
00:47
00:40
320
02.08
01:47
01:31
01:20
01:11
01:04
00:58
00:53
00:46
Table 3 Partial UVB Time Chart
The UVA Time Chart (not shown here) is configured the same as the UVB Time Chart. The
doses are listed in Joules and the energy output levels and times are different than the UVB
output levels. Each chart is clearly marked.
16.3.2
Setting Up a Treatment Time
1.
After unlocking device, tap the “Treat Patient” button.
2.
Touch the ▲/▼ arrows to set desired Treatment Time
3.
Tap the Enter Arrow
to enter Treatment Confirmation screen.
4.
Ensure patient has put on goggles and positioned themselves per the instructions
.
5.
Confirm Treatment Time.
•
If Treatment Time is correct, tap Enter Arrow
to begin treatment.
•
If changes need to be made to Treatment Time, tap Return Arrow
to return to
6.
Following delayed start, if enabled, treatment will begin, and screen will display
progress.
To repeat the same dose for another exposure, touch the Repeat button
Note:
The system beeps to signal the end of the treatment and displays the delivered
dose with the elapsed time.
16.3.3
Special Functions of the Integrated Dosimeter
•
To pause an active treatment, tap the Pause button
•
To resume a paused treatment, tap the Play button
•
To end or cancel an active treatment, first pause the treatment and then tap the Cancel
button
16.3.4
Special Notes
•
It is
not
possible to enter a time of 60 seconds. You must enter 1 minute instead.
•
If power goes off during a treatment, the system will remember how much of the
treatment has elapsed. When power is restored, your device provides option to resume
interrupted treatment.
→