EN-54
Technical Information
No data available
No cycle started
Cycle must be started via the Start Cycle
menu.
Display/Problem
Meaning
Possible cause and remedy
Inputs not possible as no
cycle started
No cycle started
Cycle must be started via the Start Cycle
menu.
Wake-up time is not in the
measurement time window
Wake-up time is not in
the measurement time
window
Select wake-up time within the set measure-
ment time window or change wake-up mode
to “Every day”.
If the battery discharges completely, the
date/time must be reset after the battery is
recharged and the unit is turned on.
Battery discharged due to
long measurement break,
date/time invalid, out-of-
date cycle data.
Date/time incorrect,
available cycle data
unusable
If the unit has not been used for a long time
(e.g. after a pregnancy), the battery discharg-
es and the date/time must be reset after the
battery is charged and the unit is turned on.
The out-of-date cycle data should be deleted
via cyclotest® mySoftware.