background image

                        

Shenzhen Zhiyong Electronics Co., Ltd.

 

 

 4 

leakage current from the capacitor of this device and provide anti-interference effect to the system. 

 

The right GND binding post: this binding post can be connected to the ground of the lab and safely 

release the leakage current from the capacitor of this device 

5.

 

Safety Instruction 

 

Every  phase  in  this  device  contain  large  filter  capacitor  connecting  ground  and  shell.  This  capacitor  could 

cause large leakage current and electric shock if grounding is bad. So, the three binding posts of this device 

must be well grounded. 

 

To provide better safety protection, we suggest three phase high power insulated transformer (100A per phase) 

in front. 

 

When the current of EUT is too high and cause severe heating of the LISN, extra fan is required for cooling 

process. 

 

 
 

 

 

 

 

 

 

6. Packing List 

 

Packing List 

LISN 

RF cable 

BNC(male) 2m 

BNC(female) to N(male) 

Instruction manual 

Warranty card 

Test report 

Power supply 

Input 

Insulated 

Transformer 

Filter 

                        Shielded Room 

Receiver 

EM5040D   

LISN 

 

  EUT 

CYBERTEK

Отзывы: