![CyberResearch UMDAS 0802DA Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/cyberresearch/umdas-0802da/umdas-0802da_user-manual_2703026024.webp)
-10V
+10V
Input Voltage
0
16383
Output Code
8192
Figure 3-10. Ideal ADC transfer function
The offset error is measured at mid-scale. Ideally, a zero volt input should produce an output code of 8192. Any
deviation from this is an offset error.
Figure 3-11. ADC transfer function with offset error
shows an example of a
UMDAS 0802DA
transfer function with a ±2.44 mV offset error. Offset error
affects all codes equally by shifting the entire transfer function up or down along the input voltage axis.
The accuracy plots in Figure 3-11 are drawn for clarity and are not drawn to scale.
-10V
+10V
Input Voltage
0
16383
Offset=2.44mV
Ideal
Actual
Output Code
2.44mV
8192
Gain error is a change in the slope of the transfer function from the ideal, and is typically expressed as a
percentage of full-scale. Gain error is easily converted to voltage by multiplying the full-scale (
FS
) input by the
error.
3-9
UMDAS 0802DA
User's Guid
e F
unctional Details
Содержание UMDAS 0802DA
Страница 2: ......
Страница 3: ......
Страница 4: ......
Страница 6: ...UMDAS 0802DA CyberResearch Data Acquisition iv Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 35: ......
Страница 36: ......
Страница 38: ...UMDAS 0802DA CyberResearch Data Acquisition 32 Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 40: ...UMDAS 0802DA CyberResearch Data Acquisition 34 Copyright 2006 CyberResearch Inc Intentionally Blank...
Страница 41: ......