TENSPros
Caution:
Text with a “CAUTION” indicator will explain possible
safety
infractions that could have the potential to cause minor to
moderate injury or damage to equipment.
Warning:
Text with a “WARNING” indicator will explain possible
safety
infractions that will potentially cause serious injury and
equipment damage.
Danger:
Text with a “DANGER” indicator will explain possible
safety
infractions that are imminently hazardous situations that
would
result in death or serious injury.
2.2
CAUTION
Caution:
Read, understand, and practice the precautionary and operating
instructions. Know the limitations and hazards associated with using
any ultrasound device. Observe the precautionary and operational
decals placed on the unit.
Keep yourself informed of the contraindications.
Never leave the patient unattended during treatment.
Do not operate the device when connected to any other medical
devices.
Do not operate this unit in an environment where other devices are
used that intentionally radiates electromagnetic energy in an
unshielded manner.
Ultrasound should be routinely checked before each use to determine
that all controls function normally, especially that the intensity control
does properly adjust the intensity of the ultrasonic power output in
stable manner. Also, determine that the treatment time control does
actually terminate ultrasonic power output when the timer reaches zero.
Do not use sharp objects such as pencil point or ballpoint pen to
operate the buttons on the control panel.
Before each use, inspect Ultrasound Applicator for cracks, which may
allow the ingress of conductive fluid.
Inspect Applicator cables and associated connectors before each use.
The ultrasound therapy controls unit is not designed to prevent the
ingress of water or liquids. Ingress of water of liquids could cause
malfunction of internal components of system and therefore create
risk of injury to the patient.
Caution should be used for patients with suspected or diagnosed
epilepsy.
Refer to recent surgical procedures when muscle contraction may
disrupt the healing process.
Do not use over the menstruating or pregnant uterus, and over areas
of the skin which lack normal sensation.
4
SoundCare™ Plus Pain Relief Device