Installation
21
Cat. No. 01024821
NOTE The wire connectors must be connected to the circuit board properly. The wires must exit the plug-in
connector opposite of the raised white base of the circuit board connector. Failure to properly connect
any of the connectors will result in a malfunction of the circuit board operation.
13. Connect the other end of the power cord, with the spade terminals, to the two 2.5VAC terminals on the trans-
14. Insert the Aqua-Sensor sensor probe wire harness through the Auqa-Sensor connector opening at location #3
on the controller enclosure. See
15. Tighten the nut on the interior side of the port opening on the controller enclosure. See
16. Connect the Aqua-Sensor probe wire harness bushing to the circuit board. The Aqua-Sensor probe terminal is
labeled “Aqua-Sensor.” See Figure 26.
NOTE The 230 VAC transformer does not have 2.5 VAC connections. The Aqua-Sensor probe cannot be used.
Power to circuit
board 24V
Re
d
Black
Blue
Whit
e
Connector Plugs
POSITION SENSOR CONNECTOR
MOTOR CONNECTOR
FLOW METER CONNECTOR
AQUA-SENSOR PROBE
PHONE
24v
2.5v
Power to circuit board
2.5VAC required only for
Aqua-Sensor installations
Power cable should be
plugged to 24V
connectors—right side
Aqua-Sensor power
cable should be
plugged to 2.5VAC
connectors—left side
Figure 26.
Transformer-circuit board power connection.